Comparison of Oscillometry Devices Using Active Mechanical Test Loads

Overview
Journal ERJ Open Res
Specialty Pulmonary Medicine
Date 2019 Dec 31
PMID 31886158
Citations 29
Authors
Affiliations
Soon will be listed here.
Abstract

Noninvasiveness, low cooperation demand and the potential for detailed physiological characterisation have promoted the use of oscillometry in the assessment of lung function. However, concerns have been raised about the comparability of measurement outcomes delivered by the different oscillometry devices. The present study compares the performances of oscillometers in the measurement of mechanical test loads with and without simulated breathing. Six devices (five were commercially available and one was custom made) were tested with mechanical test loads combining resistors (), gas compliances () and a tube inertance (), to mimic respiratory resistance ( ) and reactance ( ) spectra encountered in clinical practice. A ventilator was used to simulate breathing at tidal volumes of 300 and 700 mL at frequencies of 30 and 15 min, respectively. Measurements were evaluated in terms of , , , resonance frequency ( ), reactance area ( ) and resistance change between 5 and 20 or 19 Hz ( ). Increasing test loads caused progressive deviations in and from calculated values at various degrees in the different oscillometers. While mean values of were recovered acceptably, some devices exhibited serious distortions in the frequency dependences of and , leading to large errors in , , , and . The results were largely independent of the simulated breathing. Simplistic calibration procedures and mouthpiece corrections, in addition to unknown instrumental and signal processing factors, may be responsible for the large differences in oscillometry measures. Rigorous testing and ongoing harmonisation efforts are necessary to better exploit the diagnostic and scientific potential of oscillometry.

Citing Articles

Oscillometry reference values for children and adolescents.

Valach C, Wouters E, Ofenheimer A, Pompilio P, Gobbi A, Breyer M ERJ Open Res. 2024; 10(6).

PMID: 39698065 PMC: 11655022. DOI: 10.1183/23120541.00278-2024.


Small airway dysfunction measured by impulse oscillometry is associated with exacerbations and poor symptom control in patients with asthma treated in a tertiary hospital subspecialist airways disease clinic.

Beinart D, Goh E, Boardman G, Chung L Front Allergy. 2024; 5:1403894.

PMID: 39210978 PMC: 11358081. DOI: 10.3389/falgy.2024.1403894.


Small airways in asthma: Pathophysiology, identification and management.

Toumpanakis D, Usmani O Chin Med J Pulm Crit Care Med. 2024; 1(3):171-180.

PMID: 39171124 PMC: 11332871. DOI: 10.1016/j.pccm.2023.07.002.


Fetal and Early-Life Stove and Fuel Interventions and Respiratory Oscillometry Lung Function Measures in Preschool-Aged Children from Ghana.

McCollum E, Checkley W, Balakrishnan K, Clasen T, Peel J Am J Respir Crit Care Med. 2024; 209(6):625-626.

PMID: 38290112 PMC: 10945070. DOI: 10.1164/rccm.202311-2197ED.


Real-World Application of Oscillometry: Taking the LEAD.

Kaminsky D Am J Respir Crit Care Med. 2024; 209(4):356-357.

PMID: 38190700 PMC: 10878378. DOI: 10.1164/rccm.202311-2127ED.


References
1.
Bates J, Irvin C, Farre R, Hantos Z . Oscillation mechanics of the respiratory system. Compr Physiol. 2013; 1(3):1233-72. DOI: 10.1002/cphy.c100058. View

2.
Hantos Z, Czovek D, Gyurkovits Z, Szabo H, Maar B, Radics B . Assessment of respiratory mechanics with forced oscillations in healthy newborns. Pediatr Pulmonol. 2014; 50(4):344-52. DOI: 10.1002/ppul.23103. View

3.
Janssens J, Nguyen M, Herrmann F, Michel J . Diagostic value of respiratory impedance measurements in elderly subjects. Respir Med. 2001; 95(5):415-22. DOI: 10.1053/rmed.2001.1062. View

4.
Ducharme F, Jroundi I, Jean G, Lavoie Boutin G, Lawson C, Vinet B . Interdevice agreement in respiratory resistance values by oscillometry in asthmatic children. ERJ Open Res. 2019; 5(1). PMC: 6421362. DOI: 10.1183/23120541.00138-2018. View

5.
Czovek D, Shackleton C, Hantos Z, Taylor K, Kumar A, Chacko A . Tidal changes in respiratory resistance are sensitive indicators of airway obstruction in children. Thorax. 2016; 71(10):907-15. DOI: 10.1136/thoraxjnl-2015-208182. View