Abstract

The impulse oscillation system (IOS) has been developed recently to measure respiratory system resistance (Rrs) and reactance (Xrs) at different frequencies up to > or = 25 Hz. IOS has, however, not been validated against established techniques. This study compared IOS with the classical pseudorandom noise forced oscillation technique (FOT) and body plethysmographic airway resistance (Raw) in 49 subjects with a variety of lung disorders and a wide range of Raw (0.10-1.28 kPa x L(-1) x s). Rrs,IOS was slightly greater than Rrs,FOT, especially at lower frequencies, with a mean +/- SD difference at 5-6 Hz of 0.14 +/- 0.09 kPa x L(-1) x s. Comparisons with the wave-tube technique applied on two analogues indicated an overestimation by IOS. Xrs,IOS and Xrs,FOT were very similar, with a slightly higher resonant frequency with IOS than with FOT (mean difference +/- SD 1.35 +/- 3.40 Hz). Raw was only moderately correlated with Rrn,FOT and Rrs-IOS; although the mean differences were small (0.04 +/- 0.14 kPa x L(-1)s for Rrs6,FOT and -0.10 +/- 0.14 kPa x L(-1) x s for Rrs5,IOS), IOS and FOT markedly underestimated high resistance values. In conclusion, the impulse oscillation system yields respiratory system resistance and reactance values similar, but not identical to those provided by the forced oscillation technique.

Citing Articles

Oscillometry in Asthma: Respiratory Modeling and Analysis in Occupational and Work-Exacerbated Phenotypes.

Pinto M, Ribeiro C, Morisco de Sa P, Castro H, Bartholo T, Lopes A J Asthma Allergy. 2024; 17:983-1000.

PMID: 39411425 PMC: 11476752. DOI: 10.2147/JAA.S473639.


Differences between oscillometry measurements obtained by MostGraph-01 and MasterScreen-IOS in patients with asthma.

Harada S, Harada N, Sasano H, Tanabe Y, Kotajima M, Sato Y PLoS One. 2024; 19(9):e0309981.

PMID: 39240983 PMC: 11379245. DOI: 10.1371/journal.pone.0309981.


Oscillometry of the respiratory system in Parkinson's disease: physiological changes and diagnostic use.

Caldas B, Ribeiro F, Pereira J, Souza W, Lopes A, de Melo P BMC Pulm Med. 2023; 23(1):406.

PMID: 37884922 PMC: 10605979. DOI: 10.1186/s12890-023-02716-w.


Relation of impulse oscillometry and spirometry with quantitative thorax computed tomography after COVID-19 pneumonia.

Sahin M, Gokcek A, Satar S, Ergun P Rev Assoc Med Bras (1992). 2023; 69(5):e20221427.

PMID: 37222321 PMC: 10204846. DOI: 10.1590/1806-9282.20221427.


Assessing the prognostic value of respiratory oscillometry in patients with difficult-to-treat asthma.

Shen Y, Hsieh Y, Huang Y, Peng Y, Chen L, Dai F Sci Rep. 2023; 13(1):2457.

PMID: 36774404 PMC: 9922248. DOI: 10.1038/s41598-023-29672-z.