» Articles » PMID: 31574947

In Vivo Simulation of Magnesium Degradability Using a New Fluid Dynamic Bench Testing Approach

Overview
Journal Int J Mol Sci
Publisher MDPI
Date 2019 Oct 3
PMID 31574947
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

The degradation rate of magnesium (Mg) alloys is a key parameter to develop Mg-based biomaterials and ensure in vivo-mechanical stability as well as to minimize hydrogen gas production, which otherwise can lead to adverse effects in clinical applications. However, in vitro and in vivo results of the same material often differ largely. In the present study, a dynamic test bench with several single bioreactor cells was constructed to measure the volume of hydrogen gas which evolves during magnesium degradation to indicate the degradation rate in vivo. Degradation medium comparable with human blood plasma was used to simulate body fluids. The media was pumped through the different bioreactor cells under a constant flow rate and 37 °C to simulate physiological conditions. A total of three different Mg groups were successively tested: Mg WE43, and two different WE43 plasma electrolytically oxidized (PEO) variants. The results were compared with other methods to detect magnesium degradation (pH, potentiodynamic polarization (PDP), cytocompatibility, SEM (scanning electron microscopy)). The non-ceramized specimens showed the highest degradation rates and vast standard deviations. In contrast, the two PEO samples demonstrated reduced degradation rates with diminished standard deviation. The pH values showed above-average constant levels between 7.4-7.7, likely due to the constant exchange of the fluids. SEM revealed severe cracks on the surface of WE43 after degradation, whereas the ceramized surfaces showed significantly decreased signs of corrosion. PDP results confirmed the improved corrosion resistance of both PEO samples. While WE43 showed slight toxicity in vitro, satisfactory cytocompatibility was achieved for the PEO test samples. In summary, the dynamic test bench constructed in this study enables reliable and simple measurement of Mg degradation to simulate the in vivo environment. Furthermore, PEO treatment of magnesium is a promising method to adjust magnesium degradation.

Citing Articles

The Characterization of a Biodegradable Mg Alloy after Powder Bed Fusion with Laser Beam/Metal Processing for Custom Shaped Implants.

Raducanu D, Cojocaru V, Nocivin A, Drob S, Hendea R, Stanciu D Materials (Basel). 2024; 17(7).

PMID: 38612195 PMC: 11012606. DOI: 10.3390/ma17071682.


Comparison of Time-Zero Primary Stability Between a Biodegradable Magnesium Bone Staple and Metal Bone Staples for Knee Ligament Fixation: A Biomechanical Study in a Porcine Model.

Deichsel A, Glasbrenner J, Raschke M, Klimek M, Peez C, Briese T Orthop J Sports Med. 2024; 12(3):23259671241236783.

PMID: 38532766 PMC: 10964459. DOI: 10.1177/23259671241236783.


How surface-to-volume ratio affects degradation of magnesium: and studies.

Sun J, Liu S, Zou D, He X, Shi Z, Li W RSC Adv. 2024; 14(10):6805-6814.

PMID: 38405068 PMC: 10887483. DOI: 10.1039/d3ra08927d.


Interference screws manufactured from magnesium display similar primary stability for soft tissue anterior cruciate ligament graft fixation compared to a biocomposite material - a biomechanical study.

Deichsel A, Glasbrenner J, Raschke M, Klimek M, Peez C, Briese T J Exp Orthop. 2023; 10(1):103.

PMID: 37815666 PMC: 10564698. DOI: 10.1186/s40634-023-00663-3.


Newly Developed Resorbable Magnesium Biomaterials for Orbital Floor Reconstruction in Caprine and Ovine Animal Models-A Prototype Design and Proof-of-Principle Study.

Tomic J, Wiederstein-Grasser I, Schanbacher M, Weinberg A J Funct Biomater. 2023; 14(7).

PMID: 37504834 PMC: 10381438. DOI: 10.3390/jfb14070339.


References
1.
Schneider L, Korber A, Grabbe S, Dissemond J . Influence of pH on wound-healing: a new perspective for wound-therapy?. Arch Dermatol Res. 2006; 298(9):413-20. DOI: 10.1007/s00403-006-0713-x. View

2.
Ge Z, Yang F, Goh J, Ramakrishna S, Lee E . Biomaterials and scaffolds for ligament tissue engineering. J Biomed Mater Res A. 2006; 77(3):639-52. DOI: 10.1002/jbm.a.30578. View

3.
Jahnen-Dechent W, Ketteler M . Magnesium basics. Clin Kidney J. 2015; 5(Suppl 1):i3-i14. PMC: 4455825. DOI: 10.1093/ndtplus/sfr163. View

4.
Reifenrath J, Marten A, Angrisani N, Eifler R, Weizbauer A . In vitro and in vivo corrosion of the novel magnesium alloy Mg-La-Nd-Zr: influence of the measurement technique and in vivo implant location. Biomed Mater. 2015; 10(4):045021. DOI: 10.1088/1748-6041/10/4/045021. View

5.
Hornberger H, Virtanen S, Boccaccini A . Biomedical coatings on magnesium alloys - a review. Acta Biomater. 2012; 8(7):2442-55. DOI: 10.1016/j.actbio.2012.04.012. View