Cycled Pulsing to Mitigate Thermal Damage for Multi-electrode Irreversible Electroporation Therapy

Overview
Publisher Informa Healthcare
Specialties Oncology
Pharmacology
Date 2019 Sep 24
PMID 31542973
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

This study evaluates the effects of various pulsing paradigms, on the irreversible electroporation (IRE) lesion, induced electric current, and temperature changes using a perfused porcine liver model. A 4-monopolar electrode array delivered IRE therapy varying the pulse length and inter-pulse delay to six porcine mechanically perfused livers. Pulse paradigms included six forms of cycled pulsing schemes and the conventional pulsing scheme. Finite element models provided further insight into the effects of cycled pulsing on the temperature and thermal injury distribution. 'Single pulse cycle with no interpulse delay' deposited maximum average energy (2.34 ± 0.35 kJ) and produced the largest ratio of thermally damaged tissue area and IRE ablation area from all other pulse schemes (18.22%   8.11,  < .0001 all pairwise comparisons). These compared favorably to the conventional algorithm (2.09 ± 0.37 kJ, 3.49%   2.20,  < .0001, all comparisons). Though no statistical significance was found between groups, the '5 pulse cycle, 0 s delay' pulse paradigm produced the largest average IRE ablation cross sectional area (11.81 ± 1.97 cm), while conventional paradigm yielded an average of 8.90 ± 0.91 cm. Finite element modeling indicated a '10 pulse cycle, 10 s delay' generated the least thermal tissue damage and '1 pulse cycle, 0 s delay' pulse cycle sequence the most (0.47 vs. 3.76 cm), over a lengthier treatment time (16.5 vs. 6.67 minutes). Subdividing IRE pulses and adding delays throughout the treatment can reduce white tissue coagulation and electric current, while maintaining IRE treatment sizes.

Citing Articles

Temperature Distribution on Classical Two Needles IRE Setup Versus a Single Needle Prototype.

Jouni A, Baragona M, Pedersoli F, Ritter A Technol Cancer Res Treat. 2024; 23:15330338241288342.

PMID: 39440388 PMC: 11500228. DOI: 10.1177/15330338241288342.


A pilot clinical assessment of biphasic asymmetric pulsed field ablation catheter for pulmonary vein isolation.

Chen B, Lv C, Cui Y, Lu C, Cai H, Xue Z Front Cardiovasc Med. 2024; 11:1266195.

PMID: 38385135 PMC: 10879394. DOI: 10.3389/fcvm.2024.1266195.


Harnessing the Electrochemical Effects of Electroporation-Based Therapies to Enhance Anti-tumor Immune Responses.

Salameh Z, Aycock K, Alinezhadbalalami N, Imran K, McKillop I, Allen I Ann Biomed Eng. 2023; 52(1):48-56.

PMID: 37989902 PMC: 10781785. DOI: 10.1007/s10439-023-03403-x.


Real-Time Temperature Rise Estimation during Irreversible Electroporation Treatment through State-Space Modeling.

Campelo S, Jacobs 4th E, Aycock K, Davalos R Bioengineering (Basel). 2022; 9(10).

PMID: 36290467 PMC: 9598795. DOI: 10.3390/bioengineering9100499.


A Comparative Modeling Study of Thermal Mitigation Strategies in Irreversible Electroporation Treatments.

Aycock K, Campelo S, Davalos R J Heat Transfer. 2022; 144(3):031206.

PMID: 35833151 PMC: 8823459. DOI: 10.1115/1.4053199.