Abstract

Purpose: To evaluate a scintillator detector for patient-specific quality assurance of VMAT radiosurgery plans.

Methods: The detector was comprised of a 1 mm diameter, 1 mm high scintillator coupled to an acrylic optical fiber. Sixty VMAT SRS plans for treatment of single targets having sizes ranging from 3 mm to 30.2 mm equivalent diameter (median 16.3 mm) were selected. The plans were delivered to a 20 cm × 20 cm x 15 cm water equivalent plastic phantom having either the scintillator detector or radiochromic film at the center. Calibration films were obtained for each measurement session. The films were scanned and converted to dose using a 3-channel technique.

Results: The mean difference between scintillator and film was -0.45% (95% confidence interval -0.1% to 0.8%). For target equivalent diameter smaller than the median, the mean difference was 1.1% (95% confidence interval 0.5% to 1.7%). For targets larger than the median, the mean difference was -0.2% (95% confidence interval -0.7% to 0.1%).

Conclusions: The scintillator detector response is independent of target size for targets as small as 3 mm and is well-suited for patient-specific quality assurance of VMAT SRS plans. Further work is needed to evaluate the accuracy for VMAT plans that treat multiple targets using a single isocenter.

Citing Articles

Feasibility of determining external beam radiotherapy dose using LuSy dosimeter.

Wahabi J, Wong J, Mahdiraji G, Ung N J Appl Clin Med Phys. 2024; 25(6):e14387.

PMID: 38778567 PMC: 11163501. DOI: 10.1002/acm2.14387.


Field output correction factors of stereotactic cones for a diode detector: Dependence on cone size, measurement setup, reference field size and photon energy.

Lee Y, Kim Y J Radiosurg SBRT. 2022; 8(2):127-136.

PMID: 36275130 PMC: 9489070.


Evaluation of a two-dimensional diode array for patient-specific quality assurance of HyperArc.

Popple R, Sullivan R, Yuan Y, Wu X, Covington E J Appl Clin Med Phys. 2021; 22(12):203-210.

PMID: 34725909 PMC: 8664143. DOI: 10.1002/acm2.13438.


Quality Assurance for Small-Field VMAT SRS and Conventional-Field IMRT Using the Exradin W1 Scintillator.

Huang Z, Qiao J, Yang C, Liu M, Wang J, Han X Technol Cancer Res Treat. 2021; 20:15330338211036542.

PMID: 34328800 PMC: 8327019. DOI: 10.1177/15330338211036542.

References
1.
Ezzell G, Galvin J, Low D, Palta J, Rosen I, Sharpe M . Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee. Med Phys. 2003; 30(8):2089-115. DOI: 10.1118/1.1591194. View

2.
Beddar A, Mackie T, Attix F . Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical consideration. Phys Med Biol. 1992; 37(10):1883-900. DOI: 10.1088/0031-9155/37/10/006. View

3.
Beddar A, Mackie T, Attix F . Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements. Phys Med Biol. 1992; 37(10):1901-13. DOI: 10.1088/0031-9155/37/10/007. View

4.
Frelin A, Fontbonne J, Ban G, Colin J, Labalme M, Batalla A . Spectral discrimination of Cerenkov radiation in scintillating dosimeters. Med Phys. 2005; 32(9):3000-6. DOI: 10.1118/1.2008487. View

5.
Guillot M, Gingras L, Archambault L, Beddar S, Beaulieu L . Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: a comparison study of calibration procedures and validation in Cerenkov light-dominated situations. Med Phys. 2011; 38(4):2140-50. PMC: 3078161. DOI: 10.1118/1.3562896. View