Application of the Exradin W1 Scintillator to Determine Ediode 60017 and MicroDiamond 60019 Correction Factors for Relative Dosimetry Within Small MV and FFF Fields
Overview
Nuclear Medicine
Radiology
Affiliations
In this work we use EBT3 film measurements at 10 MV to demonstrate the suitability of the Exradin W1 (plastic scintillator) for relative dosimetry within small photon fields. We then use the Exradin W1 to measure the small field correction factors required by two other detectors: the PTW unshielded Ediode 60017 and the PTW microDiamond 60019. We consider on-axis correction-factors for small fields collimated using MLCs for four different TrueBeam energies: 6 FFF, 6 MV, 10 FFF and 10 MV. We also investigate percentage depth dose and lateral profile perturbations. In addition to high-density effects from its silicon sensitive region, the Ediode exhibited a dose-rate dependence and its known over-response to low energy scatter was found to be greater for 6 FFF than 6 MV. For clinical centres without access to a W1 scintillator, we recommend the microDiamond over the Ediode and suggest that 'limits of usability', field sizes below which a detector introduces unacceptable errors, can form a practical alternative to small-field correction factors. For a dosimetric tolerance of 2% on-axis, the microDiamond might be utilised down to 10 mm and 15 mm field sizes for 6 MV and 10 MV, respectively.
Godson H, Manickam R, Ponmalar Y, Ganesh K, Saminathan S, Chandraraj V J Med Phys. 2024; 49(1):73-83.
PMID: 38828075 PMC: 11141751. DOI: 10.4103/jmp.jmp_148_23.
Ando Y, Okada M, Matsumoto N, Ikuhiro K, Ishihara S, Kiriu H Phys Eng Sci Med. 2024; 47(3):1177-1189.
PMID: 38753285 DOI: 10.1007/s13246-024-01438-5.
Grid/lattice therapy: consideration of small field dosimetry.
Das I, Khan A, Dogan S, Longo M Br J Radiol. 2024; 97(1158):1088-1098.
PMID: 38552328 PMC: 11135801. DOI: 10.1093/bjr/tqae060.
Influence of magnetic field on a novel scintillation dosimeter in a 1.5 T MR-linac.
Oolbekkink S, van Asselen B, Woodings S, Wolthaus J, de Vries J, van Appeldoorn A J Appl Clin Med Phys. 2023; 25(1):e14180.
PMID: 38011008 PMC: 10795437. DOI: 10.1002/acm2.14180.
Characteristics of a plastic scintillation detector in photon beam dosimetry.
Das I, Sohn J, Lim S, Sengupta B, Feijoo M, Yadav P J Appl Clin Med Phys. 2023; 25(1):e14209.
PMID: 37983685 PMC: 10795454. DOI: 10.1002/acm2.14209.