Cardiac I-MIBG Normal Uptake Values Are Population-specific: Results from a Cohort of Controls over 60 Years of Age
Overview
Affiliations
Purpose: Cardiac I-MIBG image interpretation is affected by population differences and technical factors. We recruited older adults without cognitive decline and compared their cardiac MIBG uptake with results from the literature.
Methods: Phantom calibration confirmed that cardiac uptake results from Japan could be applied to our center. We recruited 31 controls, 17 individuals with dementia with Lewy bodies (DLB) and 15 with Alzheimer's disease (AD). Images were acquired 20 minutes and four hours after injection using Siemens cameras with medium-energy low-penetration (MELP) collimators. Local normal heart-to-mediastinum (HMR) ratios were compared to Japanese results.
Results: Siemens gamma cameras with MELP collimators should give HMRs very close to the calibrated values used in Japan. However, our cut-offs with controls were lower at 2.07 for early and 1.86 for delayed images. Applying our lower cut-off to the dementia patients may increase the specificity of cardiac MIBG imaging for DLB diagnosis in a UK population without reducing sensitivity.
Conclusions: Our local HMR cut-off values are lower than in Japan, higher than in a large US study but similar to those found in another UK center. UK centers using other cameras and collimators may need to use different cut-offs to apply our results.
Longitudinal changes in cardiac mIBG scintigraphy in mild cognitive impairment with Lewy bodies.
Greenfinch G, Hamilton C, Donaghy P, Firbank M, Barnett N, Allan L BJPsych Open. 2024; 10(6):e223.
PMID: 39635755 PMC: 11698149. DOI: 10.1192/bjo.2024.766.
Nakajima K, Matsumura T, Komatsu J, Wakabayashi H, Ono K, Kinuya S Ann Nucl Med. 2024; 38(10):814-824.
PMID: 38869809 PMC: 11401792. DOI: 10.1007/s12149-024-01950-4.
Okuda K, Nakajima K, Kitamura C, Ljungberg M, Hosoya T, Kirihara Y J Nucl Cardiol. 2023; 30(4):1630-1641.
PMID: 36740650 PMC: 10372132. DOI: 10.1007/s12350-023-03198-3.
Thomas A, Hamilton C, Heslegrave A, Barker S, Durcan R, Lawley S Mov Disord. 2022; 37(7):1495-1504.
PMID: 35318733 PMC: 9540809. DOI: 10.1002/mds.28994.
Accuracy of Cardiac Innervation Scintigraphy for Mild Cognitive Impairment With Lewy Bodies.
Roberts G, Durcan R, Donaghy P, Lawley S, Ciafone J, Hamilton C Neurology. 2021; 96(23):e2801-e2811.
PMID: 33883238 PMC: 8205462. DOI: 10.1212/WNL.0000000000012060.