» Articles » PMID: 31471076

The Utility of IPhone Oximetry Apps: A Comparison with Standard Pulse Oximetry Measurement in the Emergency Department

Overview
Journal Am J Emerg Med
Specialty Emergency Medicine
Date 2019 Sep 1
PMID 31471076
Citations 17
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: To determine if a correlation exists between 3 iphone pulse ox applications' measurements and the standard pulse oximetry (SpO2) and whether these applications can accurately determine hypoxia.

Methods: Three applications reportedly measuring SpO2 were downloaded onto an iPhone 5s. Two of these applications used the onboard light and camera lens "Pulse Oximeter" (Pox) and "Heart Rate and Pulse Oximeter" (Ox) and one used an external device that plugged into the iphone (iOx). Patients in the ED were enrolled with chief complaints of cardiac/pulmonary origin or a SpO2 ≤ 94%. All measurements were compared to controls. Concordance correlation coefficients, sensitivity, and specificity were calculated.

Results: A total of 191 patients were enrolled. The concordance correlation of iOx with control was 0.55 (CI 0.46, 0.63), POx was 0.01 (CI -0.09, 0.11), and Ox was 0.07 (CI -0.02, 0.15). 68/191 patients (35%) were found to have hypoxemia. Sensitivities for detecting hypoxia were 69%, 0%, and 7% for iOx, POx, and Ox, respectively. Specificities were 89%, 100%, and 89%. Even iOx (the most accurate) 21 (11%) were incorrectly classified nonhypoxic, and 22 (12%) were incorrectly classified hypoxic.

Conclusions: While iOx has modest concordance with control, Ox and POx showed almost none. The iOx device was best in correctly identifying hypoxia patients, but almost 1/4 of patients were incorrectly classified. The three apps provided inaccurate SpO2 measurements and had limited to no ability to accurately detect hypoxia. These apps should not be relied upon to provide accurate SpO2 measurements in emergent, even austere conditions.

Citing Articles

Families' perceptions of consumer-grade, inexpensive oxygen saturation monitors.

Charland D, Kovesi T Pediatr Pulmonol. 2024; 59(12):3349-3354.

PMID: 39109914 PMC: 11601012. DOI: 10.1002/ppul.27203.


Equity-Driven Sensing System for Measuring Skin Tone-Calibrated Peripheral Blood Oxygen Saturation (OptoBeat): Development, Design, and Evaluation Study.

Adams A, Mandel I, Gao Y, Heckman B, Nandakumar R, Choudhury T JMIR Biomed Eng. 2024; 7(1):e34934.

PMID: 38875699 PMC: 11041433. DOI: 10.2196/34934.


The Accuracy of Readily Available Consumer-Grade Oxygen Saturation Monitors in Pediatric Patients.

Kovesi T, Saban J, Haddad J, Reddy D, Webster R, Udupa S Respir Care. 2024; 69(4):387-394.

PMID: 38164568 PMC: 11108112. DOI: 10.4187/respcare.11258.


Reliability of smartphone measurements of peripheral oxygen saturation and heart rate in hypotensive patients measurement of vital signs with smartphones.

Totuk A, Bayramoglu B, Tayfur I Heliyon. 2023; 9(2):e13145.

PMID: 36814605 PMC: 9939538. DOI: 10.1016/j.heliyon.2023.e13145.


Smartphone camera oximetry in an induced hypoxemia study.

Hoffman J, Viswanath V, Tian C, Ding X, Thompson M, Larson E NPJ Digit Med. 2022; 5(1):146.

PMID: 36123367 PMC: 9483471. DOI: 10.1038/s41746-022-00665-y.


References
1.
Tomlinson S, Behrmann S, Cranford J, Louie M, Hashikawa A . Accuracy of Smartphone-Based Pulse Oximetry Compared with Hospital-Grade Pulse Oximetry in Healthy Children. Telemed J E Health. 2017; 24(7):527-535. DOI: 10.1089/tmj.2017.0166. View

2.
Wilson B, Cowan H, Lord J, Zuege D, Zygun D . The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study. BMC Emerg Med. 2010; 10:9. PMC: 2876142. DOI: 10.1186/1471-227X-10-9. View

3.
Seifi S, Khatony A, Moradi G, Abdi A, Najafi F . Accuracy of pulse oximetry in detection of oxygen saturation in patients admitted to the intensive care unit of heart surgery: comparison of finger, toe, forehead and earlobe probes. BMC Nurs. 2018; 17:15. PMC: 5905124. DOI: 10.1186/s12912-018-0283-1. View

4.
Bilan N, Behbahan A, Abdinia B, Mahallei M . Validity of pulse oximetry in detection of hypoxaemia in children: comparison of ear, thumb and toe probe placements. East Mediterr Health J. 2010; 16(2):218-22. View

5.
Berkenbosch J, Tobias J . Comparison of a new forehead reflectance pulse oximeter sensor with a conventional digit sensor in pediatric patients. Respir Care. 2006; 51(7):726-31. View