Local DRLs and Automated Risk Estimation in Paediatric Interventional Cardiology
Overview
Authors
Affiliations
Introduction: Cardiac catheterization procedures result in high radiation doses and often multiple procedures are necessary for congenital heart disease patients. However, diagnostic reference levels (DRL) remain scarce. Our first goal was finding the optimal DRL parameter and determining appropriate DRLs. The second goal was to calculate organ doses (OD), effective doses (ED) and lifetime attributable risks (LAR) per procedure and to provide conversion factors based on dose area product (DAP).
Materials And Methods: DRLs are calculated for each procedure type, as the 75th percentile of the cumulative value per procedure from the corresponding parameter. All irradiation events in the DICOM Structured Reports were automatically processed and simulated using PCXMC, resulting in OD, ED and LAR. Using a Kruskal Wallis H test and subsequent pairwise comparisons, differences in median values of the DRL parameter between procedure types were assessed.
Results: Linear regression showed a strong correlation and narrow confidence interval between DAP and product of body weight and fluoroscopy time (BWxFT), even when all procedures (diagnostic and interventional) are combined. Only 15% of the pairwise comparisons were statistically significant for DAP normalized to BWxFT (DAPBWxFT). The latter pairs contained less frequent procedure types with significant outliers. For DAP normalized to BW (DAPBW), 38% of the pairwise comparisons showed statistically significant differences. Conversion factors from DAPBW to OD and ED were reported for various weight groups, due to the higher correlation between DAPBW and both OD and ED than between DAP and both OD and ED.
Conclusions: The P75 of DAPBWxFT for all procedures combined serves as an appropriate DRL value. This facilitates local DRL determination in smaller paediatric centres, which often have insufficient data to produce appropriate DRLs for different procedure types. Conversion factors are more reliable starting from DAPBW instead of DAP and should be used according to the appropriate BW group.
SHOULD PATIENT AGE AND BODY MASS BE TAKEN IN CONSIDERATION WHEN ZERO FLUORO ABLATION IS SCHEDULED?.
Krmek N, Kornyei L, Kralik I, Delic-Brkljacic D, Milosevic M, Rode M Acta Clin Croat. 2025; 63(1):142-148.
PMID: 39959323 PMC: 11827388. DOI: 10.20471/acc.2024.63.01.17.
Krmek N, Kornyei L, Kralik I, Delic-Brkljacic D, Milosevic M, Rode M Pediatr Cardiol. 2024; 45(4):804-813.
PMID: 38411709 DOI: 10.1007/s00246-024-03428-6.
Buytaert D, Vandekerckhove K, Panzer J, Campens L, Bacher K, De Wolf D Front Cardiovasc Med. 2024; 10:1292039.
PMID: 38274314 PMC: 10808650. DOI: 10.3389/fcvm.2023.1292039.
Local Diagnostic Reference Levels for Pediatric Interventional Cardiology Procedures in Argentina.
Azcurra P, Leyton F, Lucini V, Rivarola M, Trentacoste L, Marques A Children (Basel). 2023; 10(12).
PMID: 38136079 PMC: 10741962. DOI: 10.3390/children10121877.
Ubeda C, Vano E, Perez M, Jimenez P, van Deventer E, Ramirez R Children (Basel). 2023; 10(12).
PMID: 38136060 PMC: 10742238. DOI: 10.3390/children10121858.