Stressing the Accuracy: Wrist-worn Wearable Sensor Validation over Different Conditions

Overview
Specialty Psychiatry
Date 2019 Jul 24
PMID 31332802
Citations 48
Authors
Affiliations
Soon will be listed here.
Abstract

Wearable sensors are promising instruments for conducting both laboratory and ambulatory research in psychophysiology. However, scholars should be aware of their measurement error and the conditions in which accuracy is achieved. This study aimed to assess the accuracy of a wearable sensor designed for research purposes, the E4 wristband (Empatica, Milan, Italy), in measuring heart rate (HR), heart rate variability (HRV), and skin conductance (SC) over five laboratory conditions widely used in stress reactivity research (seated rest, paced breathing, orthostatic, Stroop, speech task) and two ecological conditions (slow walking, keyboard typing). Forty healthy participants concurrently wore the wristband and two gold standard measurement systems (i.e., electrocardiography and finger SC sensor). The wristband accuracy was determined by evaluating the signal quality and the correlations with and the Bland-Altman plots against gold standard-derived measurements. Moreover, exploratory analyses were performed to assess predictors of measurement error. Mean HR measures showed the best accuracy over all conditions. HRV measures showed satisfactory accuracy in seated rest, paced breathing, and recovery conditions but not in dynamic conditions, including speaking. Accuracy was diminished by wrist movements, cognitive and emotional stress, nonstationarity, and larger wrist circumferences. Wrist SC measures showed neither correlation nor visual resemblance with finger SC signal, suggesting that the two sites may reflect different phenomena. Future studies are needed to assess the responsivity of wrist SC to emotional and cognitive stress. Limitations and implications for laboratory and ambulatory research are discussed.

Citing Articles

The Feasibility of Ambulatory Heart Rate Variability Monitoring in Non-Suicidal Self-Injury.

van Hoorn A, Mankee-Williams A, Lewis G, Mellili R, Eccles J, Ottaviani C Healthc Technol Lett. 2025; 12(1):e70007.

PMID: 40027108 PMC: 11868733. DOI: 10.1049/htl2.70007.


Quality in Question: Assessing the Accuracy of Four Heart Rate Wearables and the Implications for Psychophysiological Research.

Sinichi M, Gevonden M, Krabbendam L Psychophysiology. 2025; 62(2):e70004.

PMID: 39905563 PMC: 11794680. DOI: 10.1111/psyp.70004.


Selecting optimal wearables for measuring physiological arousal in robot-delivered mindfulness-based exercises.

Vacaru S, Lau L, Frederiks K, Sterkenburg P, Barakova E Adv Robot. 2024; 38(19-20):1364-1377.

PMID: 39507502 PMC: 11537300. DOI: 10.1080/01691864.2024.2369797.


Longitudinal single-subject neuroimaging study reveals effects of daily environmental, physiological, and lifestyle factors on functional brain connectivity.

Triana A, Salmi J, Hayward N, Saramaki J, Glerean E PLoS Biol. 2024; 22(10):e3002797.

PMID: 39378200 PMC: 11460715. DOI: 10.1371/journal.pbio.3002797.


Exploring emotions beyond the laboratory: A review of emotional and physiological ecological momentary assessment methods in children and youth.

Sosa-Hernandez L, Vogel N, Frankiewicz K, Reaume C, Drew A, McVey Neufeld S Psychophysiology. 2024; 61(12):e14699.

PMID: 39367539 PMC: 11579238. DOI: 10.1111/psyp.14699.