Repeatability and Reproducibility of Multiparametric Magnetic Resonance Imaging of the Liver

Abstract

As the burden of liver disease reaches epidemic levels, there is a high unmet medical need to develop robust, accurate and reproducible non-invasive methods to quantify liver tissue characteristics for use in clinical development and ultimately in clinical practice. This prospective cross-sectional study systematically examines the repeatability and reproducibility of iron-corrected T1 (cT1), T2*, and hepatic proton density fat fraction (PDFF) quantification with multiparametric MRI across different field strengths, scanner manufacturers and models. 61 adult participants with mixed liver disease aetiology and those without any history of liver disease underwent multiparametric MRI on combinations of 5 scanner models from two manufacturers (Siemens and Philips) at different field strengths (1.5T and 3T). We report high repeatability and reproducibility across different field strengths, manufacturers, and scanner models in standardized cT1 (repeatability CoV: 1.7%, bias -7.5ms, 95% LoA of -53.6 ms to 38.5 ms; reproducibility CoV 3.3%, bias 6.5 ms, 95% LoA of -76.3 to 89.2 ms) and T2* (repeatability CoV: 5.5%, bias -0.18 ms, 95% LoA -5.41 to 5.05 ms; reproducibility CoV 6.6%, bias -1.7 ms, 95% LoA -6.61 to 3.15 ms) in human measurements. PDFF repeatability (0.8%) and reproducibility (0.75%) coefficients showed high precision of this metric. Similar precision was observed in phantom measurements. Inspection of the ICC model indicated that most of the variance in cT1 could be accounted for by study participants (ICC = 0.91), with minimal contribution from technical differences. We demonstrate that multiparametric MRI is a non-invasive, repeatable and reproducible method for quantifying liver tissue characteristics across manufacturers (Philips and Siemens) and field strengths (1.5T and 3T).

Citing Articles

Practical approach to quantitative liver and pancreas MRI in children.

Kemp J, Ghosh A, Dillman J, Krishnasarma R, Manhard M, Tipirneni-Sajja A Pediatr Radiol. 2025; 55(1):36-57.

PMID: 39760887 DOI: 10.1007/s00247-024-06133-x.


The Glutamate-Serine-Glycine Index as a Biomarker to Monitor the Effects of Bariatric Surgery on Non-alcoholic Fatty Liver Disease.

Tan N, Shumbayawonda E, Cheng L, Low A, Lim C, Eng A J ASEAN Fed Endocr Soc. 2024; 39(2):54-60.

PMID: 39620184 PMC: 11604366. DOI: 10.15605/jafes.039.02.20.


Liver magnetic resonance imaging, non-alcoholic fatty liver disease and metabolic syndrome risk in pre-pubertal Mexican boys.

de Celis Alonso B, Shumbayawonda E, Beyer C, Hidalgo-Tobon S, Lopez-Martinez B, Dies-Suarez P Sci Rep. 2024; 14(1):26104.

PMID: 39478096 PMC: 11526175. DOI: 10.1038/s41598-024-77307-8.


Repeatability, Reproducibility, and Observer Variability of Cortical T1 Mapping for Renal Tissue Characterization.

Nowak M, Henningsson M, Davis T, Chowdhury N, Dennis A, Fernandes C J Magn Reson Imaging. 2024; 61(4):1914-1922.

PMID: 39468402 PMC: 11896918. DOI: 10.1002/jmri.29602.


Reference Range of Quantitative MRI Metrics Corrected T1 and Liver Fat Content in Children and Young Adults: Pooled Participant Analysis.

Shumbayawonda E, Beyer C, de Celis Alonso B, Hidalgo-Tobon S, Lopez-Martinez B, Klunder-Klunder M Children (Basel). 2024; 11(10).

PMID: 39457195 PMC: 11506660. DOI: 10.3390/children11101230.


References
1.
Hines C, Yu H, Shimakawa A, McKenzie C, Brittain J, Reeder S . T1 independent, T2* corrected MRI with accurate spectral modeling for quantification of fat: validation in a fat-water-SPIO phantom. J Magn Reson Imaging. 2009; 30(5):1215-22. PMC: 3205952. DOI: 10.1002/jmri.21957. View

2.
Younossi Z, Koenig A, Abdelatif D, Fazel Y, Henry L, Wymer M . Global epidemiology of nonalcoholic fatty liver disease-Meta-analytic assessment of prevalence, incidence, and outcomes. Hepatology. 2015; 64(1):73-84. DOI: 10.1002/hep.28431. View

3.
Reeder S, Hu H, Sirlin C . Proton density fat-fraction: a standardized MR-based biomarker of tissue fat concentration. J Magn Reson Imaging. 2012; 36(5):1011-4. PMC: 4779595. DOI: 10.1002/jmri.23741. View

4.
Bernal-Rusiel J, Greve D, Reuter M, Fischl B, Sabuncu M . Statistical analysis of longitudinal neuroimage data with Linear Mixed Effects models. Neuroimage. 2012; 66:249-60. PMC: 3586747. DOI: 10.1016/j.neuroimage.2012.10.065. View

5.
Tunnicliffe E, Banerjee R, Pavlides M, Neubauer S, Robson M . A model for hepatic fibrosis: the competing effects of cell loss and iron on shortened modified Look-Locker inversion recovery T (shMOLLI-T ) in the liver. J Magn Reson Imaging. 2016; 45(2):450-462. DOI: 10.1002/jmri.25392. View