Influence of Novel Design Alteration of Pedicle Screw on Pull-out Strength: A Finite Element Study
Overview
Authors
Affiliations
Background: We conducted a finite element study to assess the effectiveness of a novel pedicle screw design with two alterations in the distal and proximal portions.
Methods: Finite element (FE) models of 24 vertebrae were constructed using computed tomographic data. Pull-out strength of 4 different pedicle screws were compared. The basic screw design was a dual threaded one (PS0), in which the proximal portion is double-threaded (cortical thread), and the distal portion is single-threaded (cancellous thread). In PS1, the inter-thread double-core shape was added to PS0 in the distal portion. Compared to PS0, in PS2, the proximal portion was elongated by 5 mm. PS3 had both PS1 and PS2 features. In addition, the 24 vertebrae were classified into 3 groups based on volumetric bone mineral density (vBMD) of the vertebral body: low <120 mg/cm, moderate 120-170 mg/cm, and high >170 mg/cm.
Results: The mean pull-out strengths (±SD) were 1137 ± 500 N, 1188 ± 520 N, 1191 ± 512 N, and 1242 ± 538 N for PS0, PS1, PS2, and PS3, respectively. In PS1, there was significant difference in the incremental ratio of pull-out strength to PS0 between the low and high vBMD groups (3.7 ± 1.6% vs. 5.0 ± 1.0%, p = 0.006). In PS2, there was a significant difference in the incremental ratio to PS0 between the moderate and high vBMD groups (7.6 ± 4.0% vs. 3.3 ± 1.8%, p < 0.001). In PS3, there was a significant difference in the incremental ratio to PS0 between the moderate and high vBMD groups (12.1 ± 4.8% vs. 8.5 ± 2.1%, p = 0.003).
Conclusions: The two design alterations showed the combined additive effect in the PS3 design. The moderate vBMD group has a balanced bone property to reflect the combined effects of the PS1 and PS2 design alterations.
Rouyin A, Einafshar M, Arjmand N J Orthop Surg Res. 2024; 19(1):732.
PMID: 39506782 PMC: 11542241. DOI: 10.1186/s13018-024-05169-x.
Impact of screw tip design on screw anchorage: mechanical testing and numerical simulation.
Xie S, Cui L, Liu J, Qing P, Li J J Orthop Surg Res. 2024; 19(1):447.
PMID: 39080768 PMC: 11287948. DOI: 10.1186/s13018-024-04841-6.
Jiang J, Zhao C, Zhang J, Xu R, Chen L World J Clin Cases. 2024; 12(20):4108-4120.
PMID: 39015896 PMC: 11235545. DOI: 10.12998/wjcc.v12.i20.4108.
Yang J, Luo L, Liu J, Wang N, Xi Z, Li J Front Bioeng Biotechnol. 2024; 11:1282512.
PMID: 38260754 PMC: 10800439. DOI: 10.3389/fbioe.2023.1282512.
Wang J, Geng Z, Ma X, Zhang Z, Miao J Front Bioeng Biotechnol. 2023; 11:1263751.
PMID: 38026854 PMC: 10664026. DOI: 10.3389/fbioe.2023.1263751.