Radiofrequency Phase Encoded Half-pulses in Simultaneous Multislice Ultrashort Echo Time Imaging

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2019 Feb 14
PMID 30756426
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To describe a simultaneous multislice (SMS) ultrashort echo time (UTE) imaging method using radiofrequency phase encoded half-pulses in combination with power independent of number of slices (PINS) inversion recovery (IR) pulses to generate multiple-slice images with short T * contrasts in less than 3 min with close to an eightfold acceleration compared with a standard 2D approach.

Theory And Methods: Radiofrequency phase encoding is applied in an SMS (N = 4) excitation scheme using "sinc" half-pulses. With the use of coil sensitivity encoding (SENSE) and controlled aliasing in parallel imaging (CAIPI) in combination with a gradient echo 2D spiral readout trajectory and IR PINS pulses for contrast enhancement a fast UTE sequence is developed. Images are obtained using a model-based reconstruction method. Sequence details and performance tests on phantoms as well as the heads of healthy volunteers at 3T are presented.

Results: An SMS UTE sequence with an undersampling factor of 4 is developed using radiofrequency phase encoded half-pulses and PINS IR pulses which enables the acquisition of 8 slices at 0.86 mm resolution in less than 3-min scan time. UTE images of the head are obtained showing highlighted signal of cortical bone. Image quality and T contrast are comparable to the one obtained by corresponding single slice acquisitions with only minor deviations.

Conclusions: The method combining radiofrequency phase encoded SMS half-pulses and PINS IR pulses presents a suitable approach to SMS UTE imaging. The usage of coil sensitivity information and increased sampling density by means of interleaved slice group acquisitions allows to reduce the total scan time by a factor close to 8.

References
1.
Madio D, Lowe I . Ultra-fast imaging using low flip angles and FIDs. Magn Reson Med. 1995; 34(4):525-9. DOI: 10.1002/mrm.1910340407. View

2.
Qian Y, Williams A, Chu C, Boada F . High-resolution ultrashort echo time (UTE) imaging on human knee with AWSOS sequence at 3.0 T. J Magn Reson Imaging. 2011; 35(1):204-10. PMC: 5823004. DOI: 10.1002/jmri.22639. View

3.
Larson P, Gurney P, Nayak K, Gold G, Pauly J, Nishimura D . Designing long-T2 suppression pulses for ultrashort echo time imaging. Magn Reson Med. 2006; 56(1):94-103. PMC: 2942755. DOI: 10.1002/mrm.20926. View

4.
Robson P, Grant A, Madhuranthakam A, Lattanzi R, Sodickson D, McKenzie C . Comprehensive quantification of signal-to-noise ratio and g-factor for image-based and k-space-based parallel imaging reconstructions. Magn Reson Med. 2008; 60(4):895-907. PMC: 2838249. DOI: 10.1002/mrm.21728. View

5.
Chang E, Du J, Chung C . UTE imaging in the musculoskeletal system. J Magn Reson Imaging. 2014; 41(4):870-83. PMC: 4297256. DOI: 10.1002/jmri.24713. View