Minimally Invasive, Stereotactic, Wireless, Percutaneous Pedicle Screw Placement in the Lumbar Spine: Accuracy Rates With 182 Consecutive Screws
Overview
Affiliations
Background: Standard fluoroscopic navigation and stereotactic computed tomography-guided lumbar pedicle screw instrumentation traditionally relied on the placement of Kirshner wires (K-wires) to ensure accurate screw placement. The use of K-wires, however, is associated with a risk of morbidity due to potential ventral displacement into the retroperitoneum. We report our experience using a computer image-guided, wireless method for pedicle screw placement. We hypothesize that minimally invasive, wireless pedicle screw placement is as accurate and safe as the traditional technique using K-wires while decreasing operative time and avoiding potential complications associated with K-wires.
Methods: We conducted a retrospective review of 42 consecutive patients who underwent a stereotactic-guided, wireless lumbar pedicle screw placement. All screws were placed to provide fixation to a variety of interbody fusion constructs including anterior lumbar interbody fusion, lateral interbody fusion, and transforaminal lumbar interbody fusion. The procedures were performed using the O-arm intraoperative imaging system with StealthStation navigation (Medtronic, Memphis, TN) and Medtronic navigated instrumentation. After placing a percutaneous navigation frame into the posterior superior iliac spine or onto an adjacent spinous process, an intraoperative O-arm image was obtained to allow subsequent StealthStation navigation. Para-median incisions were selected to allow precise percutaneous access to the target pedicles. The pedicles were cannulated using either a stereotactic drill or a novel awl-tipped tap along with a low-speed/high-torque power driver. The initial trajectory into the pedicle was recorded on the Medtronic StealthStation prior to removal of the drill or awl-tap, creating a "virtual" K-wire rather than inserting an actual K-wire to allow subsequent tapping and screw insertion. Accurate screw placement is achieved by following the virtual path as an exact computer-aided design model of the screw traversing the pedicle is projected onto the display and by using audible and tactile feedback. A second O-arm scan was obtained to confirm accuracy of screw placement.
Results: A total of 20 women and 22 men (average age = 56 years) underwent a total of 182 pedicle screw placements using the stereotactic, wireless technique. The total breach rate was 9.9%, with a clinically significant breach rate of 0% (defined as >2 mm medial breach or >4 mm lateral breach) and a clinical complication rate of 0%.
Conclusions: Wireless, percutaneous placement of lumbar pedicle screws using computed tomography-guided stereotactic navigation is a safe, reproducible technique with very high accuracy rates.
Development of Augmented Reality Vision for Osteosynthesis Using a 3D Camera.
Morita J, Ikumi A, Nakatani T, Noguchi H, Mishima H, Ishii T Cureus. 2024; 16(5):e60479.
PMID: 38882985 PMC: 11180535. DOI: 10.7759/cureus.60479.
Pereira B, ONeill L, Sawa A, Zhou J, Wangsawatwong P, Lehrman J Int J Spine Surg. 2023; 17(4):484-491.
PMID: 37076254 PMC: 10478698. DOI: 10.14444/8470.
Singhatanadgige W, Pholprajug P, Songthong K, Yingsakmongkol W, Triganjananun C, Kotheeranurak V Int J Spine Surg. 2022; 16(1):151-158.
PMID: 35177525 PMC: 9519077. DOI: 10.14444/8183.
Workflow and Efficiency of Robotic-Assisted Navigation in Spine Surgery.
Avrumova F, Sivaganesan A, Alluri R, Vaishnav A, Qureshi S, Lebl D HSS J. 2021; 17(3):302-307.
PMID: 34539271 PMC: 8436346. DOI: 10.1177/15563316211026658.
Manson N, El-Mughayyar D, Bigney E, Richardson E, Abraham E Adv Orthop. 2020; 2020:7949216.
PMID: 32802519 PMC: 7415111. DOI: 10.1155/2020/7949216.