» Articles » PMID: 30350613

Calibration Using a Single-Point External Reference Material Harmonizes Quantitative Mass Spectrometry Proteomics Data Between Platforms and Laboratories

Overview
Journal Anal Chem
Specialty Chemistry
Date 2018 Oct 24
PMID 30350613
Citations 16
Authors
Affiliations
Soon will be listed here.
Abstract

Mass spectrometry (MS) measurements are not inherently calibrated. Researchers use various calibration methods to assign meaning to arbitrary signal intensities and improve precision. Internal calibration (IC) methods use internal standards (IS) such as synthesized or recombinant proteins or peptides to calibrate MS measurements by comparing endogenous analyte signal to the signal from known IS concentrations spiked into the same sample. However, recent work suggests that using IS as IC introduces quantitative biases that affect comparison across studies because of the inability of IS to capture all sources of variation present throughout an MS workflow. Here, we describe a single-point external calibration strategy to calibrate signal intensity measurements to a common reference material, placing MS measurements on the same scale and harmonizing signal intensities between instruments, acquisition methods, and sites. We demonstrate data harmonization between laboratories and methodologies using this generalizable approach.

Citing Articles

A framework for quality control in quantitative proteomics.

Tsantilas K, Merrihew G, Robbins J, Johnson R, Park J, Plubell D bioRxiv. 2024; .

PMID: 38645098 PMC: 11030400. DOI: 10.1101/2024.04.12.589318.


In utero pulse injection of isotopic amino acids quantifies protein turnover rates during murine fetal development.

Baeza J, Coons B, Lin Z, Riley J, Mendoza M, Peranteau W Cell Rep Methods. 2024; 4(2):100713.

PMID: 38412836 PMC: 10921036. DOI: 10.1016/j.crmeth.2024.100713.


Multiple reaction monitoring assays for large-scale quantitation of proteins from 20 mouse organs and tissues.

Michaud S, Petrosova H, Sinclair N, Kinnear A, Jackson A, McGuire J Commun Biol. 2024; 7(1):6.

PMID: 38168632 PMC: 10762018. DOI: 10.1038/s42003-023-05687-0.


A Multipathway Phosphopeptide Standard for Rapid Phosphoproteomics Assay Development.

Searle B, Chien A, Koller A, Hawke D, Herren A, Kim J Mol Cell Proteomics. 2023; 22(10):100639.

PMID: 37657519 PMC: 10561125. DOI: 10.1016/j.mcpro.2023.100639.


2019 Association of Biomolecular Resource Facilities Multi-Laboratory Data-Independent Acquisition Proteomics Study.

Kirkpatrick J, Stemmer P, Searle B, Herring L, Martin L, Midha M J Biomol Tech. 2023; 34(2).

PMID: 37435391 PMC: 10332336. DOI: 10.7171/3fc1f5fe.9b78d780.


References
1.
Hoofnagle A, Whiteaker J, Carr S, Kuhn E, Liu T, Massoni S . Recommendations for the Generation, Quantification, Storage, and Handling of Peptides Used for Mass Spectrometry-Based Assays. Clin Chem. 2016; 62(1):48-69. PMC: 4830481. DOI: 10.1373/clinchem.2015.250563. View

2.
Geiger T, Cox J, Ostasiewicz P, Wisniewski J, Mann M . Super-SILAC mix for quantitative proteomics of human tumor tissue. Nat Methods. 2010; 7(5):383-5. DOI: 10.1038/nmeth.1446. View

3.
Ghaemmaghami S, Huh W, Bower K, Howson R, Belle A, Dephoure N . Global analysis of protein expression in yeast. Nature. 2003; 425(6959):737-41. DOI: 10.1038/nature02046. View

4.
Agger S, Marney L, Hoofnagle A . Simultaneous quantification of apolipoprotein A-I and apolipoprotein B by liquid-chromatography-multiple- reaction-monitoring mass spectrometry. Clin Chem. 2010; 56(12):1804-13. PMC: 3103773. DOI: 10.1373/clinchem.2010.152264. View

5.
Craig H . Isotopic Variations in Meteoric Waters. Science. 1961; 133(3465):1702-3. DOI: 10.1126/science.133.3465.1702. View