Performance Evaluation and Measurement Uncertainty Determination of the New Version of the Abbott ARCHITECT 25-OH Vitamin D 5P02 Assay

Overview
Specialty Pathology
Date 2018 Oct 12
PMID 30307462
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: We evaluated performance and measurement uncertainty of the newly released ARCHITECT 25-OH vitamin D 5P02 assay (Abbott Diagnostics, Abbott Park, IL).

Methods: In total, 300 samples were used to compare the results from ELECSYS Vitamin D Total (Roche Diagnostics, Mannheim, Germany) and ADVIA Centaur Vitamin D Total (Siemens, Tarrytown, NY). To quantify the measurement uncertainty, 25 samples and four levels of standard reference material (SRM) were measured using isotope dilution-liquid chromatography-tandem mass spectrometry according to international guidelines.

Results: The results of the ARCHITECT assay were equivalent to other immunoassays, but correlation coefficients were lower than the recommended criterion. SRM level 2 was considered adequate for uncertainty estimation, and the expanded measurement uncertainty of the ARCHITECT assay was 4.2%, which was superior to the other two assays.

Conclusions: The restandardized ARCHITECT assay has acceptable performance in a clinical setting. However, there is still a need for further standardization of total vitamin D measurement among the automated immunoassays.

Citing Articles

Recent Information on Vitamin D Deficiency in an Adult Korean Population Visiting Local Clinics and Hospitals.

Choi R, Cho S, Lee S, Lee E Nutrients. 2022; 14(9).

PMID: 35565944 PMC: 9101007. DOI: 10.3390/nu14091978.


Interlaboratory comparison of 25-hydroxyvitamin D assays: Vitamin D Standardization Program (VDSP) Intercomparison Study 2 - Part 2 ligand binding assays - impact of 25-hydroxyvitamin D and 24R,25-dihydroxyvitamin D on assay performance.

Wise S, Camara J, Burdette C, Hahm G, Nalin F, Kuszak A Anal Bioanal Chem. 2021; 414(1):351-366.

PMID: 34435207 PMC: 8748290. DOI: 10.1007/s00216-021-03577-0.


Interlaboratory comparison of 25-hydroxyvitamin D assays: Vitamin D Standardization Program (VDSP) Intercomparison Study 2 - Part 1 liquid chromatography - tandem mass spectrometry (LC-MS/MS) assays - impact of 3-epi-25-hydroxyvitamin D on assay....

Wise S, Camara J, Burdette C, Hahm G, Nalin F, Kuszak A Anal Bioanal Chem. 2021; 414(1):333-349.

PMID: 34432104 DOI: 10.1007/s00216-021-03576-1.


Vitamin D testing: advantages and limits of the current assays.

Altieri B, Cavalier E, Bhattoa H, Perez-Lopez F, Lopez-Baena M, Perez-Roncero G Eur J Clin Nutr. 2020; 74(2):231-247.

PMID: 31907366 DOI: 10.1038/s41430-019-0553-3.