Monte Carlo Dose Verification of VMAT Treatment Plans Using Elekta Agility 160-leaf MLC

Overview
Journal Phys Med
Date 2018 Oct 4
PMID 30278982
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

In this study, we verified volumetric modulated arc therapy (VMAT) plans in an Elekta Synergy system with an integrated Agility 160-leaf multileaf collimator (MLC) by comparing them with Monte Carlo (MC)-calculated dose distributions using the AAPM TG-119 structure sets. The head configuration of the linear accelerator with the integrated MLC was simulated with the EGSnrc/BEAMnrc code. Firstly, the dosimetric properties of the MLC were evaluated with the MC technique and film measurements. Next, VMAT plans were created with the Pinnacle treatment planning system (TPS) for four regions in the AAPM TG-119 structures. They were then verified by comparing them with MC-calculated dose distributions using dose volume histograms (DVHs) and three-dimensional (3D) gamma analysis. The MC simulations for the Agility MLC dosimetric properties were in acceptable agreement with measurements. TPS-VMAT plans using TG-119 structure sets agreed with MC dose distributions within 2% in the comparison of D in planning target volumes (PTVs) evaluated from DVHs. In contrast, higher dose regions such as D, D, and D in PTVs for TPS tended to be smaller than MC values. This tendency was particularly noticeable for mock head and neck with complicated structures. In 3D gamma analysis, the passing rates with 3%/3mm criteria in PTVs were ≥99%, except for mock head and neck (89.5%). All passing rates for organs at risk (OARs) were in acceptable agreement of >96%. It is useful to verify dose distributions of PTVs and OARs in TPS-VMAT plans by using MC dose calculations and 3D gamma analysis.

Citing Articles

Improving animal-specific radiotherapy quality assurance for kilovoltage X-ray radiotherapy using a 3D printed dog skull water phantom.

Tanabe Y, Iseri T, Onizuka R, Ishida T, Eto H, Nakaichi M Open Vet J. 2023; 13(4):427-432.

PMID: 37251269 PMC: 10219815. DOI: 10.5455/OVJ.2023.v13.i4.4.


Determination of the Small-Field Output Factor for 6 MV Photon Beam Using EGSnrc Monte Carlo.

Chuah K, Abdul Aziz M, Jayamani J J Med Phys. 2023; 47(3):301-308.

PMID: 36684700 PMC: 9846995. DOI: 10.4103/jmp.jmp_40_22.


Monte Carlo modeling of the Elekta Versa HD and patient dose calculation with EGSnrc/BEAMnrc.

Paschal H, Kabat C, Papaconstadopoulos P, Kirby N, Myers P, Wagner T J Appl Clin Med Phys. 2022; 23(9):e13715.

PMID: 35985698 PMC: 9512349. DOI: 10.1002/acm2.13715.