» Articles » PMID: 30278481

Feasibility and Efficacy of Transient Elastography Using the XL Probe to Diagnose Liver Fibrosis and Cirrhosis: A Meta-analysis

Overview
Specialty General Medicine
Date 2018 Oct 4
PMID 30278481
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Transient elastography (TE) has been validated as an effective noninvasive tool for the assessment of liver fibrosis. The XL probe is a new probe that was initially designed for use in patients with obesity. A meta-analysis was performed to assess the feasibility and efficacy of TE using the XL probe.

Methods: In September 2016, we systematically searched the PubMed and Science Direct search engines. The feasibility of TE was evaluated based on the failure rate and the results of the unreliable liver stiffness measurement (LSM). The efficacy of TE was measured using sensitivity, specificity, and summary receiver-operating characteristic as measures/indices assessed in different stages of fibrosis. Heterogeneity was measured using the chi-squared test and the Q-statistic. We used the 95% confidence interval (95% CI) as an effect measure.

Results: We included 8 studies in the meta-analysis. When the XL was compared to the M probe, the former showed a lower risk of failure rate [relative risk (RR) 0.24, 95% CI 0.14-0.38]. In patients with a body mass index ≥30 kg/m, the XL probe showed a statistically significantly lower risk of failure rate (RR 0.16, 95% CI 0.08-0.32) but no significant improvement (RR 0.76, 95% CI 0.50-1.16) in the unreliable LSM result. In patients showing liver fibrosis stage ≥F2, the XL probe showed a sensitivity of 0.56 (95% CI 0.39-0.72), specificity of 0.71 (95% CI 0.61-0.79), and an area under the curve (AUC) of 0.71. The results observed in patients with liver fibrosis stage F4 were more promising with a sensitivity of 0.84 (95% CI 0.76-0.90), specificity of 0.78 (95% CI 0.70-0.84), and an AUC of 0.88.

Conclusion: TE using the XL probe demonstrates significant diagnostic utility in patients with liver fibrosis and is likely to be more reliable than the M probe in patients with obesity. Large prospective multicenter studies are, however, necessary to establish the new cut-off values to be used for the XL probe in patients with obesity.

Citing Articles

Non-invasive diagnosis of non-alcoholic fatty liver disease: Current status and future perspective.

Wang J, Jiang S, Hu A, Zhou A, Hu T, Li H Heliyon. 2024; 10(5):e27325.

PMID: 38449611 PMC: 10915413. DOI: 10.1016/j.heliyon.2024.e27325.


Non-Invasive Measurement of Hepatic Fibrosis by Transient Elastography: A Narrative Review.

Rinaldi L, Giorgione C, Mormone A, Esposito F, Rinaldi M, Berretta M Viruses. 2023; 15(8).

PMID: 37632072 PMC: 10459581. DOI: 10.3390/v15081730.


Liver Involvement in Patients with Systemic Sclerosis: Role of Transient Elastography in the Assessment of Hepatic Fibrosis and Steatosis.

Cuomo G, Iandoli C, Galiero R, Caturano A, Di Vico C, Perretta D Diagnostics (Basel). 2023; 13(10).

PMID: 37238250 PMC: 10217307. DOI: 10.3390/diagnostics13101766.


Defining the optimal technique for endoscopic ultrasound shear wave elastography: a combined benchtop and animal model study with comparison to transabdominal shear wave elastography.

Wang T, Ryou M Clin Endosc. 2023; 56(2):229-238.

PMID: 36849118 PMC: 10073853. DOI: 10.5946/ce.2022.135.


How to Identify Advanced Fibrosis in Adult Patients with Non-Alcoholic Fatty Liver Disease (NAFLD) and Non-Alcoholic Steatohepatitis (NASH) Using Ultrasound Elastography-A Review of the Literature and Proposed Multistep Approach.

Taru M, Neamti L, Taru V, Procopciuc L, Procopet B, Lupsor-Platon M Diagnostics (Basel). 2023; 13(4).

PMID: 36832276 PMC: 9955630. DOI: 10.3390/diagnostics13040788.


References
1.
Castera L, Vergniol J, Foucher J, Le Bail B, Chanteloup E, Haaser M . Prospective comparison of transient elastography, Fibrotest, APRI, and liver biopsy for the assessment of fibrosis in chronic hepatitis C. Gastroenterology. 2005; 128(2):343-50. DOI: 10.1053/j.gastro.2004.11.018. View

2.
Lok A, McMahon B . Chronic hepatitis B. Hepatology. 2007; 45(2):507-39. DOI: 10.1002/hep.21513. View

3.
Yoneda M, Thomas E, Sclair S, Grant T, Schiff E . Supersonic Shear Imaging and Transient Elastography With the XL Probe Accurately Detect Fibrosis in Overweight or Obese Patients With Chronic Liver Disease. Clin Gastroenterol Hepatol. 2015; 13(8):1502-9.e5. DOI: 10.1016/j.cgh.2015.03.014. View

4.
Fraquelli M, Rigamonti C, Casazza G, Conte D, Donato M, Ronchi G . Reproducibility of transient elastography in the evaluation of liver fibrosis in patients with chronic liver disease. Gut. 2007; 56(7):968-73. PMC: 1994385. DOI: 10.1136/gut.2006.111302. View

5.