In-vitro Validation of 4D Flow MRI Measurements with an Experimental Pulsatile Flow Model
Overview
Authors
Affiliations
Purpose: The purpose of this study was to assess the precision of four-dimensional (4D) phase-contrast magnetic resonance imaging (PCMRI) to measure mean flow and peak velocity (V) in a pulsatile flow phantom and to test its sensitivity to spatial resolution and Venc.
Material And Methods: The pulsatile flow phantom consisted of a straight tube connected to the systemic circulation of an experimental mock circulatory system. Four-dimensional-PCMR images were acquired using different spatial resolutions (minimum pixel size: 1.5×1.5×1.5mm) and velocity encoding sensitivities (up to three times V). Mean flow and V calculated from 4D-PCMRI were compared respectively to the reference phantom flow parameters and to V obtained from two-dimensional (2D)-PCMRI.
Results: 4D-PCI measured mean flow with a precision of -0.04% to+5.46%, but slightly underestimated V when compared to 2D-PCMRI (differences ranging from -1.71% to -3.85%). 4D PCMRI mean flow measurement was influenced by spatial resolution (P<0.001) with better results obtained with smaller voxel size. There was no effect of Venc on mean flow measurement. Regarding V, neither spatial resolution nor Venc did influence the precision of the measurement.
Conclusion: Using an experimental pulsatile flow model 4D-PCMRI is accurate to measure mean flow and V with better results obtained with higher spatial resolution. We also show that Venc up to 3 times higher than V may be used with no effect on these measurements.
Oechtering T, Nowak A, Sieren M, Stroth A, Kirschke N, Wegner F J Cardiovasc Magn Reson. 2023; 25(1):22.
PMID: 36978131 PMC: 10052852. DOI: 10.1186/s12968-023-00921-4.
Soulat G, Scott M, Pathrose A, Jarvis K, Berhane H, Allen B Diagn Interv Imaging. 2022; 103(9):418-426.
PMID: 35523699 PMC: 11041270. DOI: 10.1016/j.diii.2022.04.003.
Pravdivtseva M, Gaidzik F, Berg P, Ulloa P, Larsen N, Jansen O Tomography. 2022; 8(1):457-478.
PMID: 35202203 PMC: 8880336. DOI: 10.3390/tomography8010038.
Quality Control for 4D Flow MR Imaging.
Isoda H, Fukuyama A Magn Reson Med Sci. 2022; 21(2):278-292.
PMID: 35197395 PMC: 9680545. DOI: 10.2463/mrms.rev.2021-0165.
Saitta S, Guo B, Pirola S, Menichini C, Guo D, Shan Y Front Bioeng Biotechnol. 2021; 9:742985.
PMID: 34692660 PMC: 8531216. DOI: 10.3389/fbioe.2021.742985.