» Articles » PMID: 30224854

Loop and Knot Security of a Novel Arthroscopic Sliding-locking Knot Using High-strength Sutures

Overview
Journal J Orthop
Specialty Orthopedics
Date 2018 Sep 19
PMID 30224854
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: We evaluated the loop and knot security of a novel arthroscopic knot, the Wiese knot, using different types of sutures.

Methods: The Wiese knot was tied using four different brands of braided sutures (Ethibond, Orthocord, FiberWire, and UltraBraid) with and without a series of three reversing half-hitches (RHAPs) and tested for loop and knot security.

Results: Orthocord provided the greatest amount of loop security. FiberWire delivered the highest knot security. UltraBraid had the greatest ultimate force. Three half-hitches increased the maximal load to clinical failure.

Conclusion: The biomechanical characteristics of the Wiese knot are affected by suture material qualities.

Citing Articles

Laparoscopic Surgery: A Systematic Review of Loop and Knot Security, Varying with the Suture and Sequences, Throws, Rotation and Destabilization of Half-Knots or Half-Hitches.

Koninckx P, Ussia A, Wattiez A, Kondo W, Romeo A J Clin Med. 2023; 12(19).

PMID: 37834810 PMC: 10573094. DOI: 10.3390/jcm12196166.


Biomechanics of surgical knot security: a systematic review.

Wong Y, McGrouther D Int J Surg. 2023; 109(3):481-490.

PMID: 36912691 PMC: 10389331. DOI: 10.1097/JS9.0000000000000298.


Mechanical Comparison of High-Strength Tape Suture Versus High-Strength Round Suture.

Ensminger W, McIff T, Vopat B, Mullen S, Schroeppel J Arthrosc Sports Med Rehabil. 2021; 3(5):e1525-e1534.

PMID: 34712990 PMC: 8527324. DOI: 10.1016/j.asmr.2021.07.014.


Knots Tied With High-Tensile Strength Tape Biomechanically Outperform Knots Tied With Round Suture.

Hong C, Chuang H, Hsu K, Kuan F, Chen Y, Yeh M Orthop J Sports Med. 2021; 9(10):23259671211039554.

PMID: 34671688 PMC: 8521428. DOI: 10.1177/23259671211039554.

References
1.
Lo I, Burkhart S, Chan K, Athanasiou K . Arthroscopic knots: determining the optimal balance of loop security and knot security. Arthroscopy. 2004; 20(5):489-502. DOI: 10.1016/j.arthro.2004.03.005. View

2.
Abbi G, Espinoza L, Odell T, Mahar A, Pedowitz R . Evaluation of 5 knots and 2 suture materials for arthroscopic rotator cuff repair: very strong sutures can still slip. Arthroscopy. 2006; 22(1):38-43. DOI: 10.1016/j.arthro.2005.10.010. View

3.
Barber F, Herbert M, Beavis R . Cyclic load and failure behavior of arthroscopic knots and high strength sutures. Arthroscopy. 2009; 25(2):192-9. DOI: 10.1016/j.arthro.2008.09.010. View

4.
Elkousy H, Sekiya J, Stabile K, McMahon P . A biomechanical comparison of arthroscopic sliding and sliding-locking knots. Arthroscopy. 2005; 21(2):204-10. DOI: 10.1016/j.arthro.2004.09.024. View

5.
Loutzenheiser T, Harryman 2nd D, Yung S, France M, Sidles J . Optimizing arthroscopic knots. Arthroscopy. 1995; 11(2):199-206. DOI: 10.1016/0749-8063(95)90067-5. View