Image Quality and Dose Characteristics for an O-arm Intraoperative Imaging System with Model-based Image Reconstruction
Overview
Affiliations
Purpose: To assess the imaging performance and radiation dose characteristics of the O-arm CBCT imaging system (Medtronic Inc., Littleton MA) and demonstrate the potential for improved image quality and reduced dose via model-based image reconstruction (MBIR).
Methods: Two main studies were performed to investigate previously unreported characteristics of the O-arm system. First is an investigation of dose and 3D image quality achieved with filtered back-projection (FBP) - including enhancements in geometric calibration, handling of lateral truncation and detector saturation, and incorporation of an isotropic apodization filter. Second is implementation of an MBIR algorithm based on Huber-penalized likelihood estimation (PLH) and investigation of image quality improvement at reduced dose. Each study involved measurements in quantitative phantoms as a basis for analysis of contrast-to-noise ratio and spatial resolution as well as imaging of a human cadaver to test the findings under realistic imaging conditions.
Results: View-dependent calibration of system geometry improved the accuracy of reconstruction as quantified by the full-width at half maximum of the point-spread function - from 0.80 to 0.65 mm - and yielded subtle but perceptible improvement in high-contrast detail of bone (e.g., temporal bone). Standard technique protocols for the head and body imparted absorbed dose of 16 and 18 mGy, respectively. For low-to-medium contrast (<100 HU) imaging at fixed spatial resolution (1.3 mm edge-spread function) and fixed dose (6.7 mGy), PLH improved CNR over FBP by +48% in the head and +35% in the body. Evaluation at different dose levels demonstrated 30% increase in CNR at 62% of the dose in the head and 90% increase in CNR at 50% dose in the body.
Conclusions: A variety of improvements in FBP implementation (geometric calibration, truncation and saturation effects, and isotropic apodization) offer the potential for improved image quality and reduced radiation dose on the O-arm system. Further gains are possible with MBIR, including improved soft-tissue visualization, low-dose imaging protocols, and extension to methods that naturally incorporate prior information of patient anatomy and/or surgical instrumentation.
Sommer C, Ozden I, Weyland M, Duran C, Lutters G, Scheidegger S Z Med Phys. 2022; 32(2):209-217.
PMID: 35184974 PMC: 9948851. DOI: 10.1016/j.zemedi.2022.01.001.
Three-dimensional intraoperative computed tomography imaging for zygomatic fracture repair.
Peleg O, Ianculovici C, Shuster A, Mijiritsky E, Oz I, Kleinman S J Korean Assoc Oral Maxillofac Surg. 2021; 47(5):382-387.
PMID: 34713813 PMC: 8564091. DOI: 10.5125/jkaoms.2021.47.5.382.
Zhang X, Zbijewski W, Huang Y, Uneri A, Jones C, Lo S Med Phys. 2021; 48(11):6800-6809.
PMID: 34519364 PMC: 10174643. DOI: 10.1002/mp.15221.
Sheth N, De Silva T, Uneri A, Ketcha M, Han R, Vijayan R Med Phys. 2019; 47(3):958-974.
PMID: 31863480 PMC: 8630541. DOI: 10.1002/mp.13983.
Known-component metal artifact reduction (KC-MAR) for cone-beam CT.
Uneri A, Zhang X, Yi T, Stayman J, Helm P, Osgood G Phys Med Biol. 2019; 64(16):165021.
PMID: 31287092 PMC: 7262472. DOI: 10.1088/1361-6560/ab3036.