Soft-tissue Imaging with C-arm Cone-beam CT Using Statistical Reconstruction

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2014 Feb 8
PMID 24504126
Citations 38
Authors
Affiliations
Soon will be listed here.
Abstract

The potential for statistical image reconstruction methods such as penalized-likelihood (PL) to improve C-arm cone-beam CT (CBCT) soft-tissue visualization for intraoperative imaging over conventional filtered backprojection (FBP) is assessed in this work by making a fair comparison in relation to soft-tissue performance. A prototype mobile C-arm was used to scan anthropomorphic head and abdomen phantoms as well as a cadaveric torso at doses substantially lower than typical values in diagnostic CT, and the effects of dose reduction via tube current reduction and sparse sampling were also compared. Matched spatial resolution between PL and FBP was determined by the edge spread function of low-contrast (∼ 40-80 HU) spheres in the phantoms, which were representative of soft-tissue imaging tasks. PL using the non-quadratic Huber penalty was found to substantially reduce noise relative to FBP, especially at lower spatial resolution where PL provides a contrast-to-noise ratio increase up to 1.4-2.2 × over FBP at 50% dose reduction across all objects. Comparison of sampling strategies indicates that soft-tissue imaging benefits from fully sampled acquisitions at dose above ∼ 1.7 mGy and benefits from 50% sparsity at dose below ∼ 1.0 mGy. Therefore, an appropriate sampling strategy along with the improved low-contrast visualization offered by statistical reconstruction demonstrates the potential for extending intraoperative C-arm CBCT to applications in soft-tissue interventions in neurosurgery as well as thoracic and abdominal surgeries by overcoming conventional tradeoffs in noise, spatial resolution, and dose.

Citing Articles

Deformable motion compensation in interventional cone-beam CT with a context-aware learned autofocus metric.

Huang H, Liu Y, Siewerdsen J, Lu A, Hu Y, Zbijewski W Med Phys. 2024; 51(6):4158-4180.

PMID: 38733602 PMC: 11155121. DOI: 10.1002/mp.17125.


Combining physics-based models with deep learning image synthesis and uncertainty in intraoperative cone-beam CT of the brain.

Zhang X, Sisniega A, Zbijewski W, Lee J, Jones C, Wu P Med Phys. 2023; 50(5):2607-2624.

PMID: 36906915 PMC: 10175241. DOI: 10.1002/mp.16351.


Hepatic dual-contrast CT imaging: slow triple kVp switching CT with CNN-based sinogram completion and material decomposition.

Cao W, Shapira N, Maidment A, Daerr H, Noel P J Med Imaging (Bellingham). 2022; 9(1):014003.

PMID: 35127967 PMC: 8802083. DOI: 10.1117/1.JMI.9.1.014003.


Advanced Image Registration and Reconstruction using the O-Arm System: Dose Reduction, Image Quality, and Guidance using Known-Component Models.

Uneri A, Zhang X, Stayman J, Helm P, Osgood G, Theodore N Proc SPIE Int Soc Opt Eng. 2021; 10576.

PMID: 34290469 PMC: 8292104. DOI: 10.1117/12.2293874.


Convergence criterion for MBIR based on the local noise-power spectrum: Theory and implementation in a framework for accelerated 3D image reconstruction with a morphological pyramid.

Sisniega A, Stayman J, Capostagno S, Weiss C, Ehtiati T, Siewerdsen J Proc SPIE Int Soc Opt Eng. 2021; 11072.

PMID: 34267413 PMC: 8278841. DOI: 10.1117/12.2534896.


References
1.
Nielsen T, Manzke R, Proksa R, Grass M . Cardiac cone-beam CT volume reconstruction using ART. Med Phys. 2005; 32(4):851-60. DOI: 10.1118/1.1869052. View

2.
Kolditz D, Meyer M, Kyriakou Y, Kalender W . Comparison of extended field-of-view reconstructions in C-arm flat-detector CT using patient size, shape or attenuation information. Phys Med Biol. 2010; 56(1):39-56. DOI: 10.1088/0031-9155/56/1/003. View

3.
Daly M, Siewerdsen J, Cho Y, Jaffray D, Irish J . Geometric calibration of a mobile C-arm for intraoperative cone-beam CT. Med Phys. 2008; 35(5):2124-36. PMC: 2809734. DOI: 10.1118/1.2907563. View

4.
Mueller K, Yagel R . Rapid 3-D cone-beam reconstruction with the simultaneous algebraic reconstruction technique (SART) using 2-D texture mapping hardware. IEEE Trans Med Imaging. 2001; 19(12):1227-37. DOI: 10.1109/42.897815. View

5.
Schmidgunst C, Ritter D, Lang E . Calibration model of a dual gain flat panel detector for 2D and 3D x-ray imaging. Med Phys. 2007; 34(9):3649-64. DOI: 10.1118/1.2760024. View