Independent Validation of Metric Optimized Gating for Fetal Cardiovascular Phase-contrast Flow Imaging

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2018 Aug 31
PMID 30159933
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To validate metric optimized gating phase-contrast MR (MOG PC-MR) flow measurements for a range of fetal flow velocities in phantom experiments. 2) To investigate intra- and interobserver variability for fetal flow measurements at an imaging center other than the original site.

Methods: MOG PC-MR was compared to timer/beaker measurements in a pulsatile flow phantom using a heart rate (∼145 bpm), nozzle diameter (∼6 mm), and flow range (∼130-700 mL/min) similar to fetal imaging. Fifteen healthy fetuses were included for intra- and interobserver variability in the fetal descending aorta and umbilical vein.

Results: Phantom MOG PC-MR flow bias and variability was 2% ± 23%. Accuracy of MOG PC-MR was degraded for flow profiles with low velocity-to-noise ratio. Intra- and interobserver coefficients of variation were 6% and 19%, respectively, for fetal descending aorta; and 10% and 17%, respectively, for the umbilical vein.

Conclusion: Phantom validation showed good agreement between MOG and conventionally gated PC-MR, except for cases with low velocity-to-noise ratio, which resulted in MOG misgating and underestimated peak velocities and warranted optimization of sequence parameters to individual fetal vessels. Inter- and intraobserver variability for fetal MOG PC-MR imaging were comparable to previously reported values.

Citing Articles

The onset of coarctation of the aorta before birth: Mechanistic insights from fetal arch anatomy and haemodynamics.

Hermida U, van Poppel M, Sabry M, Keramati H, Steinweg J, Simpson J Comput Biol Med. 2024; 182:109077.

PMID: 39265477 PMC: 11846778. DOI: 10.1016/j.compbiomed.2024.109077.


Fetal cardiovascular blood flow MRI: techniques and applications.

van Amerom J, Goolaub D, Schrauben E, Sun L, Macgowan C, Seed M Br J Radiol. 2022; 96(1147):20211096.

PMID: 35687661 PMC: 10321246. DOI: 10.1259/bjr.20211096.


Structured analysis of the impact of fetal motion on phase-contrast MRI flow measurements with metric optimized gating.

Schulz A, Lloyd D, van Poppel M, Roberts T, Steinweg J, Pushparajah K Sci Rep. 2022; 12(1):5395.

PMID: 35354868 PMC: 8967860. DOI: 10.1038/s41598-022-09327-1.


Quantification of blood flow in the fetus with cardiovascular magnetic resonance imaging using Doppler ultrasound gating: validation against metric optimized gating.

Ryd D, Sun L, Steding-Ehrenborg K, Bidhult S, Kording F, Ruprecht C J Cardiovasc Magn Reson. 2019; 21(1):74.

PMID: 31783877 PMC: 6883707. DOI: 10.1186/s12968-019-0586-8.


Fetal Cardiac MRI: A Review of Technical Advancements.

Roy C, van Amerom J, Marini D, Seed M, Macgowan C Top Magn Reson Imaging. 2019; 28(5):235-244.

PMID: 31592990 PMC: 6791520. DOI: 10.1097/RMR.0000000000000218.


References
1.
Heiberg E, Sjogren J, Ugander M, Carlsson M, Engblom H, Arheden H . Design and validation of Segment--freely available software for cardiovascular image analysis. BMC Med Imaging. 2010; 10:1. PMC: 2822815. DOI: 10.1186/1471-2342-10-1. View

2.
Tonge H, Struijk P, Wladimiroff J . Blood flow measurements in the fetal descending aorta: technique and clinics. Clin Cardiol. 1984; 7(6):323-9. DOI: 10.1002/clc.4960070602. View

3.
Gatehouse P, Rolf M, Graves M, Hofman M, Totman J, Werner B . Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements. J Cardiovasc Magn Reson. 2010; 12:5. PMC: 2818657. DOI: 10.1186/1532-429X-12-5. View

4.
Kok R, de Vries M, Heerschap A, van den Berg P . Absence of harmful effects of magnetic resonance exposure at 1.5 T in utero during the third trimester of pregnancy: a follow-up study. Magn Reson Imaging. 2004; 22(6):851-4. DOI: 10.1016/j.mri.2004.01.047. View

5.
Seed M, van Amerom J, Yoo S, Al Nafisi B, Grosse-Wortmann L, Jaeggi E . Feasibility of quantification of the distribution of blood flow in the normal human fetal circulation using CMR: a cross-sectional study. J Cardiovasc Magn Reson. 2012; 14:79. PMC: 3534594. DOI: 10.1186/1532-429X-14-79. View