» Articles » PMID: 29907506

Toward Planning Target Volume Margin Reduction for the Prostate Using Intrafraction Motion Correction with Online KV Imaging and Automatic Detection of Implanted Gold Seeds

Overview
Publisher Elsevier
Specialties Oncology
Radiology
Date 2018 Jun 17
PMID 29907506
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The imaging application Auto Beam Hold (ABH) allows for the online analysis of 2-dimensional kV images acquired during treatment. ABH can automatically detect fiducial markers and initiate a beam interrupt. In this study, we investigate the practical use and results of this intrafraction monitoring tool for patients with prostate cancer who have implanted gold seeds treated with a RapidArc technique.

Methods And Materials: A total of 105 patients were included. For setup, the seeds were lined up using 2 orthogonal 2-dimensional kV images. After the setup procedure, ABH was applied at an interval of 3 seconds. The software requires a maximum-allowed deviation to be defined for each seed, which is referred to as a deviation limit (DL). Online, the ABH application evaluates the position of the seeds and indicates for each seed whether or not it exceeds the DL. Patients were divided in 3 groups. For the first group ABH was used with the DL at 6 mm, which corresponds to the planning target volume (PTV) margin. For the second group, the DL was set at 5 mm with an unchanged PTV margin of 6 mm. For the third group, the PTV margin was reduced to 5 mm with a DL of 5 mm. Offline, we performed an analysis of the number of beam stops and resulting re-setups.

Results: ABH initiated a beam interrupt 223 times (13%) during a total of 1736 sessions. By decreasing the DL from 6 mm to 5 mm, the amount of workload for re-setups increased from 6% (group 1) to 14% (groups 2 and 3). Re-setup, 3-dimensional shifts larger than the PTV margin were found in 44%, 35%, and 45% for groups 1,2, and 3, respectively.

Conclusions: Intrafraction imaging of prostate position during treatment using automatic detection of implanted gold seeds was successfully implemented. PTV margins were safely reduced from 6mm to 5mm without a substantial increase in workload.

Citing Articles

Stereotactic Body Radiation Therapy (SBRT) in prostate cancer in the presence of hip prosthesis - is it a contraindication? A narrative review.

Dube S, Pareek V, Barthwal M, Antony F, Sasaki D, Rivest R BMC Urol. 2024; 24(1):152.

PMID: 39061006 PMC: 11282858. DOI: 10.1186/s12894-024-01479-8.


[The MIRAGE Trial (MRI-guided stereotactic body radiotherapy for prostate cancer) - Precision at its best?].

Ehret F Strahlenther Onkol. 2024; 200(3):255-258.

PMID: 38182832 PMC: 10876491. DOI: 10.1007/s00066-023-02194-3.


Implementation of triggered kilovoltage imaging for stereotactic radiotherapy of the spine for patients with spinal fixation hardware.

Cetnar A, Degnan M, Pichler J, Jain S, Morelli S, Thomas E Phys Imaging Radiat Oncol. 2023; 25:100422.

PMID: 36875327 PMC: 9978845. DOI: 10.1016/j.phro.2023.100422.


Magnetic resonance imaging-guided radiotherapy for intermediate- and high-risk prostate cancer: Trade-off between planning target volume margin and online plan adaption.

Tetar S, Bruynzeel A, Verweij L, Bohoudi O, Slotman B, Rosario T Phys Imaging Radiat Oncol. 2022; 23:92-96.

PMID: 35844255 PMC: 9283928. DOI: 10.1016/j.phro.2022.06.013.


Effect of intrafraction adaptation on PTV margins for MRI guided online adaptive radiotherapy for rectal cancer.

Kensen C, Janssen T, Betgen A, Wiersema L, Peters F, Remeijer P Radiat Oncol. 2022; 17(1):110.

PMID: 35729587 PMC: 9215022. DOI: 10.1186/s13014-022-02079-2.