The Effect of Thermocycling on the Degree of Conversion and Mechanical Properties of a Microhybrid Dental Resin Composite
Overview
Authors
Affiliations
Objective: The purpose of this study was to investigate the degree of conversion (DC) and mechanical properties of a microhybrid Filtek Z250 (3M ESPE) resin composite after aging.
Method: The specimens were fabricated using circular molds to investigate Vickers microhardness (Vickers hardness number [VHN]) and DC, and were prepared according to ISO 4049 for flexural strength testing. The initial DC (%) of discs was recorded using attenuated total reflectance-Fourier transforming infrared spectroscopy. The initial VHN of the specimens was measured using a microhardness tester under a load of 300 g for 15 seconds and the flexural strength test was carried out with a universal testing machine (crosshead speed, 0.5 mm/min). The specimens were then subjected to thermocycling in 5°C and 55°C water baths. Properties were assessed after 1,000-10,000 cycles of thermocycling. The surfaces were evaluated using scanning electron microscopy (SEM). Data were analyzed using 1-way analysis of variance followed by the Tukey honest significant difference test.
Results: Statistical analysis showed that DC tended to increase up to 4,000 cycles, with no significant changes. VHN and flexural strength values significantly decreased upon thermal cycling when compared to baseline ( < 0.05). However, there was no significant difference between initial and post-thermocycling VHN results at 1,000 cycles. SEM images after aging showed deteriorative changes in the resin composite surfaces.
Conclusions: The Z250 microhybrid resin composite showed reduced surface microhardness and flexural strength and increased DC after thermocycling.
Fidan M, Cankaya N Discov Nano. 2025; 20(1):43.
PMID: 39979501 PMC: 11842679. DOI: 10.1186/s11671-025-04205-x.
Irgin C BMC Oral Health. 2024; 24(1):1533.
PMID: 39709403 PMC: 11662843. DOI: 10.1186/s12903-024-05353-9.
Saati K, Valizadeh S, Rahmaniparast A, Karimi M Front Dent. 2024; 21:35.
PMID: 39588288 PMC: 11586639. DOI: 10.18502/fid.v21i35.16473.
Effect of CAD-CAM block thickness and translucency on the polymerization of luting materials.
Dogu Kaya B, Ozturk S, Senol A, Kahramanoglu E, Yilmaz Atali P, Tarcin B BMC Oral Health. 2024; 24(1):1384.
PMID: 39548438 PMC: 11566408. DOI: 10.1186/s12903-024-05171-z.
Tribst J, Veerman A, Pereira G, Kleverlaan C, Dal Piva A Clin Pract. 2024; 14(5):1940-1952.
PMID: 39311304 PMC: 11417927. DOI: 10.3390/clinpract14050154.