Comparison of the Effect of Gel Used in Two Different Serum Separator Tubes for Thyroid Function Tests

Overview
Journal J Clin Lab Anal
Publisher Wiley
Date 2018 Mar 8
PMID 29512190
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Selection and verification of blood collection tubes is an important preanalytical issue in clinical laboratories. Today, gel tubes are commonly used with many advantages, although they are known to cause interference in immunoassay methods. In this study, we aimed to compare SSTs of two different suppliers (Ayset clot activator & Gel and Becton Dickinson (BD) Vacutainer SST II advance) with reference tubes and evaluate the effect of storage time in terms of commonly used endocrine tests such as thyroid-stimulating hormone (TSH), free thyroxine (fT4), and free triiodothyronine (fT3).

Methods: Fifty-five volunteers were included in the study. Samples were taken into three different tubes and analyzed for serum TSH, fT4, and fT3 on Architect ci8200 Immunoassay System. Clinical decision levels were estimated using total allowable error (TEa).

Results: No difference was found between tubes in terms of TSH, fT3, and fT4 levels. From a statistical standpoint, TSH and fT4 levels were no longer stable during 24, 48, and 72 hours storage time periods. However, their variations were not clinically significant.

Conclusion: Ayset clot activator & Gel tubes and BD Vacutainer SST II advance tubes have comparable results with glass tube in terms of TSH, fT3, and fT4 levels on Architect ci8200 Immunoassay Systems. From a clinical standpoint, serum TSH, fT4, and fT3 concentrations may be considered as stable when storing these tubes over 72 hours.

Citing Articles

Serum or Plasma (and Which Plasma), That Is the Question.

Vignoli A, Tenori L, Morsiani C, Turano P, Capri M, Luchinat C J Proteome Res. 2022; 21(4):1061-1072.

PMID: 35271285 PMC: 8981325. DOI: 10.1021/acs.jproteome.1c00935.


Comparison of the effect of gel used in two different serum separator tubes for thyroid function tests.

Ercan M, Firat Oguz E, Akbulut E, Yilmaz M, Turhan T J Clin Lab Anal. 2018; 32(6):e22427.

PMID: 29512190 PMC: 6816829. DOI: 10.1002/jcla.22427.

References
1.
Karppi J, Akerman K, Parviainen M . Suitability of collection tubes with separator gels for collecting and storing blood samples for therapeutic drug monitoring (TDM). Clin Chem Lab Med. 2000; 38(4):313-20. DOI: 10.1515/CCLM.2000.045. View

2.
Kilinc A, Duzoylum A, Uncugil C, Yucel D . Falsely increased free triiodothyronine in sera stored in serum separator tubes. Clin Chem. 2002; 48(12):2296-7. View

3.
Schouwers S, Brandt I, Willemse J, van Regenmortel N, Uyttenbroeck W, Wauters A . Influence of separator gel in Sarstedt S-Monovette® serum tubes on various therapeutic drugs, hormones, and proteins. Clin Chim Acta. 2011; 413(1-2):100-4. DOI: 10.1016/j.cca.2011.08.037. View

4.
Landt M, Smith C, Hortin G . Evaluation of evacuated blood-collection tubes: effects of three types of polymeric separators on therapeutic drug-monitoring specimens. Clin Chem. 1993; 39(8):1712-7. View

5.
Dasgupta A, Dean R, Saldana S, Kinnaman G, McLawhon R . Absorption of therapeutic drugs by barrier gels in serum separator blood collection tubes. Volume- and time-dependent reduction in total and free drug concentrations. Am J Clin Pathol. 1994; 101(4):456-61. DOI: 10.1093/ajcp/101.4.456. View