Effect of Bone Quality, Implant Design, and Surgical Technique on Primary Implant Stability

Overview
Specialty Dentistry
Date 2018 Jan 25
PMID 29363678
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Adaptation of the surgical protocol and implant selection depending on bone quality has been advocated. It was the goal of this in-vitro study to quantify the effect of implant geometry and implant site preparation on micromotion at the implant-bone interface.

Method And Materials: A parametric study was conducted with implants differing in shape, length, and diameter placed in polyurethane foam material differing in density and structure following various methods of implant site preparation. Measurement parameters included bone quality in the cervical and apical area, insertion torque, implant stability, and implant displacement under load application. Statistical analysis was based on ANOVA and Pearson's product moment correlation with the level of significance set at α = .05.

Results: Compressive testing of bone allowed different bone densities (P = .000) to be distinguished. Minor changes in bone density caused by implant site preparation with osteotomes could not be detected (P > .05). Undersized drilling (P = .001), the presence of a cortical layer (P = .000), and the use of osteotomes (P = .000) led to a significant increase in implant insertion torque. Thread cutting reduced insertion torque (P = .000) whereas the use of short (P = 1.000) and wide diameter (P = .235) implants had no effect. Implant stability measurements did not always allow for differentiation of implants varying in shape and placed with different protocols. Increasing bone density led to a general reduction in maximum implant displacement under loading. Significant correlations between all parameters were found, with the exception of residual implant displacement, which only correlated with compressive bone testing in the cortical (P = .0341) and trabecular (P = .0359) areas.

Conclusion: Compressive testing of bone seems to allow the prediction of implant performance.

Citing Articles

Clinical Effects of Photofunctionalization on Implant Stability and Marginal Bone Loss: Systematic Review and Meta-Analysis.

Lang X, Qiao B, Ge Z, Yan J, Zhang Y J Clin Med. 2022; 11(23).

PMID: 36498616 PMC: 9739233. DOI: 10.3390/jcm11237042.


Does the Modification of the Apical Geometry of a Dental Implant Affect Its Primary Stability? A Comparative Ex Vivo Study.

Staedt H, Heimes D, Lehmann K, Ottl P, Bjelopavlovic M, Wagner W Materials (Basel). 2021; 14(7).

PMID: 33915933 PMC: 8036940. DOI: 10.3390/ma14071728.


An In Vitro Evaluation of Primary Stability Values for Two Differently Designed Implants to Suit Immediate Loading in Very Soft Bone.

Krischik D, Ergin Tokgoz S, van Orten A, Friedmann A, Bilhan H Dent J (Basel). 2021; 9(1).

PMID: 33430004 PMC: 7827002. DOI: 10.3390/dj9010005.


Effects of Different Undersizing Site Preparations on Implant Stability.

Lemos B, Lopez-Jarana P, Falcao C, Rios-Carrasco B, Gil J, Rios-Santos J Int J Environ Res Public Health. 2020; 17(23).

PMID: 33276529 PMC: 7731014. DOI: 10.3390/ijerph17238965.


Evaluation of the primary stability in dental implants placed in low density bone with a new drilling technique, Osseodensification: an in vitro study.

Barbera-Millan J, Larrazabal-Moron C, Enciso-Ripoll J, Perez-Pevida E, Chavarri-Prado D, Gomez-Adrian M Med Oral Patol Oral Cir Bucal. 2020; 26(3):e361-e367.

PMID: 33037795 PMC: 8141313. DOI: 10.4317/medoral.24231.