Comparison of Chest Pain Protocols for Electrocardiography-Gated Dual-Source Cardiothoracic CT in Children and Adults: The Effect of Tube Current Saturation on Radiation Dose Reduction
Overview
Authors
Affiliations
Objective: To compare radiation doses between conventional and chest pain protocols using dual-source retrospectively electrocardiography (ECG)-gated cardiothoracic computed tomography (CT) in children and adults and assess the effect of tube current saturation on radiation dose reduction.
Materials And Methods: This study included 104 patients (16.6 ± 7.7 years, range 5-48 years) that were divided into two groups: those with and those without tube current saturation. The estimated radiation doses of retrospectively ECG-gated spiral cardiothoracic CT were compared between conventional, uniphasic, and biphasic chest pain protocols acquired with the same imaging parameters in the same patients by using paired tests. Dose reduction percentages, patient ages, volume CT dose index values, and tube current time products per rotation were compared between the two groups by using unpaired tests. A value < 0.05 was considered significant.
Results: The volume CT dose index values of the biphasic chest pain protocol (10.8 ± 3.9 mGy) were significantly lower than those of the conventional protocol (12.2 ± 4.7 mGy, < 0.001) and those of the uniphasic chest pain protocol (12.9 ± 4.9 mGy, < 0.001). The dose-saving effect of biphasic chest pain protocol was significantly less with a saturated tube current (4.5 ± 10.2%) than with unsaturated tube current method (14.8 ± 11.5%, < 0.001). In 76 patients using 100 kVp, patient age showed no significant differences between the groups with and without tube current saturation in all protocols ( > 0.05); the groups with tube current saturation showed significantly higher volume CT dose index values ( < 0.01) and tube current time product per rotation ( < 0.001) than the groups without tube current saturation in all protocols.
Conclusion: The radiation dose of dual-source retrospectively ECG-gated spiral cardiothoracic CT can be reduced by approximately 15% by using the biphasic chest pain protocol instead of the conventional protocol in children and adults if radiation dose parameters are further optimized to avoid tube current saturation.
Liu S, Hua X, Zhao Y, Mo H, Chen X, Wang W Int J Surg. 2024; 111(1):771-780.
PMID: 38905490 PMC: 11745586. DOI: 10.1097/JS9.0000000000001862.
Goo H, Park S Pediatr Cardiol. 2023; 45(1):24-31.
PMID: 38012401 DOI: 10.1007/s00246-023-03342-3.
Goo H, Park S Pediatr Radiol. 2023; 53(12):2528-2538.
PMID: 37603066 DOI: 10.1007/s00247-023-05734-2.
Goo H Pediatr Cardiol. 2022; 43(5):1104-1113.
PMID: 35107628 DOI: 10.1007/s00246-022-02830-2.
Goo H Korean J Radiol. 2021; 22(6):901-911.
PMID: 33660457 PMC: 8154780. DOI: 10.3348/kjr.2020.0787.