Investigating the Dosimetric Effects of Grid Size on Dose Calculation Accuracy Using Volumetric Modulated Arc Therapy in Spine Stereotactic Radiosurgery

Overview
Date 2018 Jan 4
PMID 29296454
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Sharp dose gradients between the target and the spinal cord are critical to achieve dose constraints in spine stereotactic radiosurgery (SRS), however the accuracy of the doses to the spinal cord at these high dose gradients is sensitive to the how the dose is sampled across the structure using a discretized isotropic calculation grid. In this study, the effect of the grid size (GS) on the dosimetric accuracy of volumetric modulated arc therapy (VMAT) spine SRS plans was investigated.

Methods: The Eclipse v11.0 Anisotropic Analytical Algorithm (AAA) algorithm was used for dose calculation. Plan qualities of fifty treatment plans were evaluated with a GS of 2.5 (AAA's default value), 1.5 and 1mm. All plans were prescribed to the 90% isodose line in 1 fraction. Parameters used for plan comparison included the distance-to-fall-off (DTF) between the 90% and 50% isodose levels in the axial plane, planning tumor volume (PTV) coverage to 99%, 95%, 5% and 0.03cc, dose to 10% (Cord_D10%) and 0.03cc (Cord_D0.03cc) of the spinal cord sub volume. The dosimetric accuracy was evaluated based on film dosimetry percent gamma pass rate, line profile through the cord. Calculation times between different grid sizes as well as DVH algorithm differences between two treatment planning systems (Eclipse vs Velocity) were compared. Paired t-test was used to investigate the statistical significance.

Results: The DTF decreased for all plans with 1mm compared to 1.5mm and 2.5mm GS (2.52±0.54mm, 2.83±0.58mm, 3.30±0.64, p<0.001). Relative to the 1mm GS, Cord_D0.03cc and Cord_D10% increased by 6.24% and 7.81% with the 1.5mm GS, and 9.80% and 13% with the 2.5mm GS. Film analysis demonstrated higher gamma pass rates for 1.5mm GS compared to 1 and 2.5mm GS (95.9%±5.4%, 94.3%±6.0%, 93.6%±5.4%, p<0.001), however 1mm GS showed better agreement in the high dose gradient near the cord. Calculation times for 1mm GS plans increased for 1.5 and 2.5mm GS (61% and 84%, p<0.001). The average difference between the two treatment planning systems was approximately 0-1.2%. A maximum difference of 5.9% occurred for Cord_D0.03cc for the 1mm GS.

Conclusion: Plans calculated with a 1mm grid size resulted in the most accurate representation of the dose delivered to the cord, however resulted in less uniform dose distributions in the high dose region of the PTV. The use of a 1.5mm grid size may balance accurate cord dose and PTV coverage, while also being more practical with respect to computation time.

Citing Articles

Dosimetric impact of arc simulation angular resolution in single-isocentre multi-target stereotactic radiosurgery.

Murugan P, Manickam R, Rajamanickam T, Muthu S, Dinesan C, Appunu K Radiol Phys Technol. 2025; 18(1):227-237.

PMID: 39812945 DOI: 10.1007/s12194-024-00876-w.


Dose prescription for stereotactic body radiotherapy: general and organ-specific consensus statement from the DEGRO/DGMP Working Group Stereotactic Radiotherapy and Radiosurgery.

Brunner T, Boda-Heggemann J, Burgy D, Corradini S, Dieckmann U, Gawish A Strahlenther Onkol. 2024; 200(9):737-750.

PMID: 38997440 PMC: 11343978. DOI: 10.1007/s00066-024-02254-2.


Evaluation of output factors of different radiotherapy planning systems using Exradin W2 plastic scintillator detector.

Ando Y, Okada M, Matsumoto N, Ikuhiro K, Ishihara S, Kiriu H Phys Eng Sci Med. 2024; 47(3):1177-1189.

PMID: 38753285 DOI: 10.1007/s13246-024-01438-5.


The dosimetric impact of titanium implants in spinal SBRT using four commercial treatment planning algorithms.

Liu C, Cho Y, Magnelli A, Angelov L, Balagamwala E, Chao S J Appl Clin Med Phys. 2023; 24(10):e14070.

PMID: 37540084 PMC: 10562029. DOI: 10.1002/acm2.14070.


Plan quality analysis of stereotactic ablative body radiotherapy treatment planning in liver tumor.

Watcharawipha A, Chakrabandhu S, Kongsa A, Tippanya D, Chitapanarux I J Appl Clin Med Phys. 2023; 24(7):e13948.

PMID: 36857202 PMC: 10338832. DOI: 10.1002/acm2.13948.


References
1.
Snyder K, Wen N, Huang Y, Kim J, Zhao B, Siddiqui S . Use of jaw tracking in intensity modulated and volumetric modulated arc radiation therapy for spine stereotactic radiosurgery. Pract Radiat Oncol. 2014; 5(3):e155-e162. DOI: 10.1016/j.prro.2014.09.002. View

2.
Yang J, Tang G, Zhang P, Hunt M, Lim S, LoSasso T . Dose calculation for hypofractionated volumetric-modulated arc therapy: approximating continuous arc delivery and tongue-and-groove modeling. J Appl Clin Med Phys. 2016; 17(2):3-13. PMC: 4831077. DOI: 10.1120/jacmp.v17i2.4989. View

3.
Ryu S, Fang Yin F, Rock J, Zhu J, Chu A, Kagan E . Image-guided and intensity-modulated radiosurgery for patients with spinal metastasis. Cancer. 2003; 97(8):2013-8. DOI: 10.1002/cncr.11296. View

4.
Wu Q, Yoo S, Kirkpatrick J, Thongphiew D, Yin F . Volumetric arc intensity-modulated therapy for spine body radiotherapy: comparison with static intensity-modulated treatment. Int J Radiat Oncol Biol Phys. 2009; 75(5):1596-604. DOI: 10.1016/j.ijrobp.2009.05.005. View

5.
Niemierko A, Goitein M . The influence of the size of the grid used for dose calculation on the accuracy of dose estimation. Med Phys. 1989; 16(2):239-47. DOI: 10.1118/1.596419. View