The Inspired Sinewave Technique: A Comparison Study With Body Plethysmography in Healthy Volunteers

Overview
Date 2017 Dec 29
PMID 29282434
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

The inspired sinewave technique is a noninvasive method to measure airway dead space, functional residual capacity, pulmonary blood flow, and lung inhomogeneity simultaneously. The purpose of this paper was to assess the repeatability and accuracy of the current device prototype in measuring functional residual capacity, and also participant comfort when using such a device. To assess within-session repeatability, six sinewave measurements were taken over two-hour period in 17 healthy volunteers. To assess day-to-day repeatability, measurements were taken over 16 days in 3 volunteers. To assess accuracy, sinewave measurements were compared to body plethysmography in 44 healthy volunteers. Finally, 18 volunteers who experienced the inspired sinewave device, body plethysmography and spirometry were asked to rate the comfort of each technique on a scale of 1-10. The repeatability coefficients for dead space, functional residual capacity, and blood flow were 48.7 ml, 0.48L, and 2.4L/min respectively. Bland-Altman analyses showed a mean BIAS(SD) of -0.68(0.42)L for functional residual capacity when compared with body plethysmography. 14 out of 18 volunteers rated the inspired sinewave device as their preferred technique. The repeatability and accuracy of functional residual capacity measurements were found to be as good as other techniques in the literature. The high level of comfort and the non-requirement of patient effort meant that, if further refined, the inspired sinewave technique could be an attractive solution for difficult patient groups such as very young children, elderly, and ventilated patients.

Citing Articles

Simulation-based optimisation to quantify heterogeneity of specific ventilation and perfusion in the lung by the Inspired Sinewave Test.

Tran M, Nguyen V, Bruce R, Crockett D, Formenti F, Phan P Sci Rep. 2021; 11(1):12627.

PMID: 34135419 PMC: 8208972. DOI: 10.1038/s41598-021-92062-w.

References
1.
Brown R, LEITH D, Enright P . Multiple breath helium dilution measurement of lung volumes in adults. Eur Respir J. 1998; 11(1):246-55. DOI: 10.1183/09031936.98.11010246. View

2.
Hahn C . Oxygen respiratory gas analysis by sine-wave measurement: a theoretical model. J Appl Physiol (1985). 1996; 81(2):985-97. DOI: 10.1152/jappl.1996.81.2.985. View

3.
Gavaghan D, Hahn C . A tidal breathing model of the forced inspired inert gas sinewave technique. Respir Physiol. 1996; 106(2):209-21. DOI: 10.1016/s0034-5687(96)00066-7. View

4.
Zwart A, Seagrave R, van Dieren A . Ventilation-perfusion ratio obtained by a noninvasive frequency response technique. J Appl Physiol. 1976; 41(3):419-24. DOI: 10.1152/jappl.1976.41.3.419. View

5.
Bland J, Altman D . Measuring agreement in method comparison studies. Stat Methods Med Res. 1999; 8(2):135-60. DOI: 10.1177/096228029900800204. View