» Articles » PMID: 28956256

Non-invasive Blood Pressure Monitoring with an Oscillometric Brachial Cuff: Impact of Arrhythmia

Overview
Publisher Springer
Date 2017 Sep 29
PMID 28956256
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Arrhythmia-induced beat-to-beat variability of blood pressure (BP) is deemed to hinder the reliability of non-invasive oscillometric measurements (NIBP) but few data support this belief. We assessed the impact of arrhythmia on a NIBP device never tested for this purpose. We compared, in intensive care unit patients with and without arrhythmia, the agreement between three pairs of NIBP (Infinity™ Delta monitor, Dräger medical systems) and invasive readings. For systolic, diastolic and mean BP, the mean bias between NIBP and invasive measurements was not higher, in 89 patients with arrhythmia, than that observed in 127 patients with regular rhythm (p = 0.93 for mean BP). Averaging three measurements overcame the higher within-subject variability of NIBP measurements during arrhythmia, and yielded similar agreement between the two techniques in patients with arrhythmia and with regular rhythm. The international organization for standardization criteria (mean bias < 5 and SD < 8 mmHg) were satisfied neither during arrhythmia nor during regular rhythm: for mean BP, mean bias of - 8.0 ± 6.5 and - 7.5 ± 6.1 mmHg, respectively. The detection of hypotension (systolic invasive BP < 90 mmHg or mean invasive BP < 65 mmHg) or hypertension (systolic invasive BP > 140 mmHg) by NIBP was similar during arrhythmia and regular rhythm [areas under the receiver operating characteristic curves (AUC) of 0.88-0.92, p > 0.13]. The detection of a 10% increase in mean invasive BP after cardiovascular intervention was also associated with similar AUCs between the two groups. Provided that triplicates are averaged, the agreement between NIBP measured with the tested device and invasive measurements was not worse during arrhythmia than during regular rhythm.

Citing Articles

Performance of inflationary oscillometric blood pressure measurement in the presence of atrial fibrillation: comparison to sinus rhythm.

Maki Y, Sakamoto Y, Abe R, Morozumi K, Toyoda D, Kotake Y J Anesth. 2025; .

PMID: 39966135 DOI: 10.1007/s00540-025-03462-5.


Can Currently Available Non-invasive Continuous Blood Pressure Monitors Replace Invasive Measurement With an Arterial Catheter?.

Gore P, Liu H, Bohringer C Cureus. 2024; 16(2):e54707.

PMID: 38529464 PMC: 10961923. DOI: 10.7759/cureus.54707.


Accuracy of noninvasive blood pressure measured at the ankle during cesarean delivery under spinal anesthesia.

Refaat S, Mostafa M, Hasanin A, Rujubali N, Fouad R, Hassabelnaby Y J Clin Monit Comput. 2020; 35(5):1211-1218.

PMID: 32860553 DOI: 10.1007/s10877-020-00583-y.


Journal of clinical monitoring and computing end of year summary 2018: hemodynamic monitoring and management.

Saugel B, Flick M, Bendjelid K, Critchley L, Vistisen S, Scheeren T J Clin Monit Comput. 2019; 33(2):211-222.

PMID: 30847738 PMC: 6420447. DOI: 10.1007/s10877-019-00297-w.


Accuracy of automated blood pressure measurements in the presence of atrial fibrillation: systematic review and meta-analysis.

Clark C, McDonagh S, McManus R J Hum Hypertens. 2019; 33(5):352-364.

PMID: 30631126 DOI: 10.1038/s41371-018-0153-z.


References
1.
Chatterjee A, DePriest K, Blair R, Bowton D, Chin R . Results of a survey of blood pressure monitoring by intensivists in critically ill patients: a preliminary study. Crit Care Med. 2010; 38(12):2335-8. DOI: 10.1097/CCM.0b013e3181fa057f. View

2.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. View

3.
Mancia G, Zanchetti A, Bohm M, Christiaens T, Cifkova R, De Backer G . 2013 ESH/ESC guidelines for the management of arterial hypertension: the Task Force for the Management of Arterial Hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC). Eur Heart J. 2013; 34(28):2159-219. DOI: 10.1093/eurheartj/eht151. View

4.
OBrien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R . European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit. 2010; 15(1):23-38. DOI: 10.1097/MBP.0b013e3283360e98. View

5.
Lip G, Kakar P, Watson T . Atrial fibrillation--the growing epidemic. Heart. 2007; 93(5):542-3. PMC: 1955544. DOI: 10.1136/hrt.2006.110791. View