Feasibility of Intraoperative Navigation for Liver Resection Using Real-time Virtual Sonography With Novel Automatic Registration System
Overview
Authors
Affiliations
Background: The clinical feasibility and usability of intraoperative ultrasonography (IOUS) tracked by computed tomography (CT) images have been proposed; however, it requires technically demanding manual registration procedure.
Study Design: A prospective study using real-time virtual sonography (RVS) with novel automatic registration system was conducted in four high-volume centers of liver resection from 2015 to 2016. The requiring time for registration of IOUS and CT images and positional error of confluence of middle hepatic venous tributaries (V8-MHV, V5-MHV) were measured in patients undergoing laparotomy.
Results: Automatic registration was successful in 43 of 52 enrolled patients (83%), with error ranges of 11.4 (3.1-69.4) mm for V8-MHV and 16.2 (4.3-66.8) mm for V5-MHV. Time required for total registration process was 36 (27-74) s.
Conclusions: The RVS with novel automatic registration system can provide quick and easy registration and acceptable accuracy, which can promote the usage of IOUS.
Surgical treatment for hepatocellular carcinoma in era of multidisciplinary strategies.
Takamoto T, Mihara Y, Nishioka Y, Ichida A, Kawaguchi Y, Akamatsu N Int J Clin Oncol. 2025; 30(3):417-426.
PMID: 39907863 PMC: 11842484. DOI: 10.1007/s10147-025-02703-7.
Sheriff N, Thomas M, Bunck A, Peterhans M, Datta R, Hellmich M Langenbecks Arch Surg. 2024; 409(1):268.
PMID: 39225933 PMC: 11371850. DOI: 10.1007/s00423-024-03456-z.
Matsubara K, Kuroda S, Kobayashi T, Ide K, Tahara H, Ohira M PLoS One. 2024; 19(6):e0306307.
PMID: 38941347 PMC: 11213349. DOI: 10.1371/journal.pone.0306307.
Takamoto T, Nara S, Ban D, Mizui T, Murase Y, Esaki M World J Surg. 2023; 47(11):2834-2845.
PMID: 37540268 DOI: 10.1007/s00268-023-07133-y.
Wu T, Huang W, He B, Guo Y, Peng G, Li M J Gastrointest Oncol. 2022; 13(3):1224-1236.
PMID: 35837192 PMC: 9274062. DOI: 10.21037/jgo-22-282.