High-pitch Dual-source CT Angiography Without ECG-gating for Imaging the Whole Aorta: Intraindividual Comparison with Standard Pitch Single-source Technique Without ECG Gating

Overview
Date 2017 Jul 14
PMID 28703104
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: We aimed to perform intraindividual comparison of computed tomography (CT) parameters, image quality, and radiation exposure between standard CT angiography (CTA) and high-pitch dual source (DS)-CTA, in subjects undergoing serial CTA of thoracoabdominal aorta.

Methods: Eighteen subjects with thoracoabdominal CTA by standard technique and high-pitch DS-CTA technique within 6 months of each other were retrieved for intraindividual comparison of image quality in thoracic and abdominal aorta. Quantitative analysis was performed by comparison of mean aortic attenuation, noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). Qualitative analysis was performed by visual assessment of motion artifacts and diagnostic confidence. Radiation exposure was quantified by effective dose. Image quality was apportioned to radiation exposure by means of figure of merit.

Results: Mean aortic attenuation and noise were higher in high-pitch DS-CTA of thoracoabdominal aorta, whereas SNR and CNR were similar in thoracic aorta and significantly lower in high-pitch DS-CTA of abdominal aorta (P = 0.024 and P = 0.016). High-pitch DS-CTA was significantly better in the first segment of thoracic aorta. Effective dose was reduced by 72% in high-pitch DS-CTA.

Conclusion: High-pitch DS-CTA without electrocardiography-gating is an effective technique for imaging aorta with very low radiation exposure and with significant reduction of motion artifacts in ascending aorta; however, the overall quality of high-pitch DS-CTA in abdominal aorta is lower than standard CTA.

Citing Articles

Progress of CT aortic angiography combined with coronary artery in the evaluation of acute aortic syndrome.

An H, Xie R, Ge Y, Wang T Front Cardiovasc Med. 2022; 9:1036982.

PMID: 36479572 PMC: 9719981. DOI: 10.3389/fcvm.2022.1036982.


Optimizing low contrast volume thoracic CT angiography: From the basics to the advanced.

Kaproth-Joslin K, Hobbs S, Rajiah P, Chaturvedi A, Chaturvedi A J Clin Imaging Sci. 2022; 12:41.

PMID: 36128360 PMC: 9479554. DOI: 10.25259/JCIS_51_2022.

References
1.
Karlo C, Leschka S, Goetti R, Feuchtner G, Desbiolles L, Stolzmann P . High-pitch dual-source CT angiography of the aortic valve-aortic root complex without ECG-synchronization. Eur Radiol. 2010; 21(1):205-12. DOI: 10.1007/s00330-010-1907-3. View

2.
Wuest W, Anders K, Schuhbaeck A, May M, Gauss S, Marwan M . Dual source multidetector CT-angiography before Transcatheter Aortic Valve Implantation (TAVI) using a high-pitch spiral acquisition mode. Eur Radiol. 2011; 22(1):51-8. DOI: 10.1007/s00330-011-2233-0. View

3.
Geleijns J, Golding S, Menzel H, Schibilla H . A workshop on quality criteria for computed tomography held in Arhus, Denmark, November 1998. Eur Radiol. 2000; 10(3):544-5. DOI: 10.1007/s003300050095. View

4.
Holmquist F, Nyman U . Eighty-peak kilovoltage 16-channel multidetector computed tomography and reduced contrast-medium doses tailored to body weight to diagnose pulmonary embolism in azotaemic patients. Eur Radiol. 2006; 16(5):1165-76. DOI: 10.1007/s00330-005-0117-x. View

5.
Bischoff B, Meinel F, Reiser M, Becker H . Novel single-source high-pitch protocol for CT angiography of the aorta: comparison to high-pitch dual-source protocol in the context of TAVI planning. Int J Cardiovasc Imaging. 2013; 29(5):1159-65. DOI: 10.1007/s10554-013-0182-1. View