Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector

Overview
Journal J Med Phys
Date 2017 Apr 14
PMID 28405102
Authors
Affiliations
Soon will be listed here.
Abstract

Advanced radiotherapy modalities such as stereotactic radiosurgery (SRS) and image-guided radiotherapy may employ very small beam apertures for accurate localized high dose to target. Accurate measurement of small radiation fields is a well-known challenge for many dosimeters. The purpose of this study was to measure total scatter factors for stereotactic cones with plastic scintillation detector and its comparison against diode detector and theoretical estimates. Measurements were performed on Novalis Tx linear accelerator for 6MV SRS beam with stereotactic cones of diameter 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm. The advantage of plastic scintillator detector is in its energy dependence. The total scatter factor was measured in water at the depth of dose maximum. Total scatter factor with plastic scintillation detector was determined by normalizing the readings to field size of 10 cm × 10 cm. To overcome energy dependence of diode detector for the determination of scatter factor with diode detector, daisy chaining method was used. The plastic scintillator detector was calibrated against the ionization chamber, and the reproducibility in the measured doses was found to be within ± 1%. Total scatter factor measured with plastic scintillation detector was 0.728 ± 0.3, 0.783 ± 0.05, 0.866 ± 0.55, 0.885 ± 0.5, and 0.910 ± 0.06 for cone sizes of 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. Total scatter factor measured with diode detector was 0.733 ± 0.03, 0.782 ± 0.02, 0.834 ± 0.07, 0.854 ± 0.02, and 0.872 ± 0.02 for cone sizes of 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. The variation in the measurement of total scatter factor with published Monte Carlo data was found to be -1.3%, 1.9%, -0.4%, and 0.4% for cone sizes of 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. We conclude that total scatter factor measurements for stereotactic cones can be adequately carried out with a plastic scintillation detector. Our results show a high level of consistency within our data and compared well with published data.

References
1.
Dieterich S, Sherouse G . Experimental comparison of seven commercial dosimetry diodes for measurement of stereotactic radiosurgery cone factors. Med Phys. 2011; 38(7):4166-73. DOI: 10.1118/1.3592647. View

2.
Guillot M, Gingras L, Archambault L, Beddar S, Beaulieu L . Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: a comparison study of calibration procedures and validation in Cerenkov light-dominated situations. Med Phys. 2011; 38(4):2140-50. PMC: 3078161. DOI: 10.1118/1.3562896. View

3.
Griessbach I, Lapp M, Bohsung J, Gademann G, Harder D . Dosimetric characteristics of a new unshielded silicon diode and its application in clinical photon and electron beams. Med Phys. 2006; 32(12):3750-4. DOI: 10.1118/1.2124547. View

4.
Ding G, Duggan D, Coffey C . Commissioning stereotactic radiosurgery beams using both experimental and theoretical methods. Phys Med Biol. 2006; 51(10):2549-66. DOI: 10.1088/0031-9155/51/10/013. View

5.
Francescon P, Beddar S, Satariano N, Das I . Variation of kQclin,Qmsr (fclin,fmsr) for the small-field dosimetric parameters percentage depth dose, tissue-maximum ratio, and off-axis ratio. Med Phys. 2014; 41(10):101708. PMC: 5175987. DOI: 10.1118/1.4895978. View