Accuracy in Blood Pressure Monitoring: The Effect of Noninvasive Blood Pressure Cuff Inflation on Intra-arterial Blood Pressure Values

Overview
Specialty Anesthesiology
Date 2017 Mar 17
PMID 28298779
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Context: Both invasive and noninvasive blood pressure (invasive arterial blood pressure [IABP] and noninvasive BP [NIBP]) monitors are used perioperatively; however, they often produce different values. The reason for this discrepancy is not clear, and it is possible that the act of cuff inflation itself might affect the IABP values, especially with the recurrent cycling of NIBP cuff.

Aim: The aim of this study was to determine the effect of ipsilateral NIBP cuff inflation on the contralateral IABP values.

Settings And Designs: Prospective, observational study.

Materials And Methods: One hundred consecutive patients were studied. The NIBP device was set to cycle every 5 min for a total of 6 times. During each cuff inflation cycle, changes in IABP values from the arterial line in the contralateral arm were recorded. A total of 582 measurements were included for data analysis.

Statistical Analysis: Chi-square, paired -test, analysis of variance.

Results: Mean (± standard deviation) changes in systolic BP (SBP), diastolic BP, and mean BP with cuff inflation were 6.7 ± 5.9, 2.6 ± 4.0, and 4.0 ± 3.9 mmHg, respectively. We observed an increase of 0-10 mmHg in SBP in majority (73.4%) of cuff inflations. The changes in IABP did not differ between the patients with or without hypertension or with the baseline SBP.

Conclusions: This study showed that there is a transient reactive rise in IABP values with NIBP cuff inflation. This is important information in the perioperative and intensive care settings, where both these measurement techniques are routinely used. The exact mechanism for this effect is not known but may be attributed to the pain and discomfort from cuff inflation.

Citing Articles

Clinical validation of a wearable ultrasound sensor of blood pressure.

Zhou S, Park G, Longardner K, Lin M, Qi B, Yang X Nat Biomed Eng. 2024; .

PMID: 39567702 DOI: 10.1038/s41551-024-01279-3.


Biceps muscle injury associated with non-invasive automatic blood pressure monitoring.

Nicol R, Lin I, Prasad L BMJ Case Rep. 2023; 16(8).

PMID: 37643817 PMC: 10465898. DOI: 10.1136/bcr-2023-256459.


Blood Pressure in Critically Ill Children: Exploratory Analyses of Concurrent Invasive and Noninvasive Measurements.

Goodwin A, Mazwi M, Somer J, Schwartz S, McEwan A, Eytan D Crit Care Explor. 2022; 3(12):e0586.

PMID: 34984339 PMC: 8718171. DOI: 10.1097/CCE.0000000000000586.


Improved agreement and diagnostic accuracy of a cuffless 24-h blood pressure measurement device in clinical practice.

Socrates T, Krisai P, Vischer A, Meienberg A, Mayr M, Burkard T Sci Rep. 2021; 11(1):1143.

PMID: 33441915 PMC: 7806663. DOI: 10.1038/s41598-020-80905-x.


The cuffless SOMNOtouch NIBP device shows poor agreement with a validated oscillometric device during 24-h ambulatory blood pressure monitoring.

Nyvad J, Christensen K, Buus N, Reinhard M J Clin Hypertens (Greenwich). 2020; 23(1):61-70.

PMID: 33350030 PMC: 8030014. DOI: 10.1111/jch.14135.


References
1.
Manios E, Vemmos K, Tsivgoulis G, Barlas G, Koroboki E, Eleni K . Comparison of noninvasive oscillometric and intra-arterial blood pressure measurements in hyperacute stroke. Blood Press Monit. 2007; 12(3):149-56. DOI: 10.1097/MBP.0b013e3280b083e2. View

2.
Takci S, Yigit S, Korkmaz A, Yurdakok M . Comparison between oscillometric and invasive blood pressure measurements in critically ill premature infants. Acta Paediatr. 2011; 101(2):132-5. DOI: 10.1111/j.1651-2227.2011.02458.x. View

3.
Charmoy A, Wurzner G, Ruffieux C, Hasler C, Cachat F, Waeber B . Reactive rise in blood pressure upon cuff inflation: cuff inflation at the arm causes a greater rise in pressure than at the wrist in hypertensive patients. Blood Press Monit. 2007; 12(5):275-80. DOI: 10.1097/MBP.0b013e3282c9ac9a. View

4.
Tomlinson L, Wilkinson I . Does it matter where we measure blood pressure?. Br J Clin Pharmacol. 2012; 74(2):241-5. PMC: 3630744. DOI: 10.1111/j.1365-2125.2012.04203.x. View

5.
Rollnik J, Schmitz N, Kugler J . Cardiovascular reactions induced by unpredictable, predictable, and controllable painful stimuli during sphygmomanometry. Int J Psychophysiol. 2001; 40(2):161-5. DOI: 10.1016/s0167-8760(00)00160-4. View