Comparison of Noninvasive Oscillometric and Intra-arterial Blood Pressure Measurements in Hyperacute Stroke

Overview
Date 2007 May 15
PMID 17496464
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: This study aims to compare automatic oscillometric blood pressure recordings with simultaneous direct intra-arterial blood pressure measurements in hyperacute stroke patients to test the accuracy of oscillometric readings.

Methods: A total of 51 first-ever stroke patients underwent simultaneous noninvasive automatic oscillometric and intra-arterial blood pressure monitoring within 3 h of ictus. Casual blood pressure was measured in both arms using a standard mercury sphygmomanometer on hospital admission. Patients who received antihypertensive medication during the blood pressure monitoring were excluded.

Results: The estimation of systolic blood pressure (SBP) using oscillometric recordings underestimated direct radial artery SBP by 9.7 mmHg (95% confidence interval: 6.5-13.0, P<0.001). In contrast, an upward bias of 5.6 mmHg (95% confidence interval: 3.5-7.7, P<0.001) was documented when noninvasive diastolic blood pressure (DBP) recordings were compared with intra-arterial DBP recordings. For SBP and DBP, the Pearson correlation coefficients between noninvasive and intra-arterial recordings were 0.854 and 0.832, respectively. When the study population was stratified according to SBP bands (group A: SBP<or=160 mmHg; group B: SBP>160 mmHg and SBP<or=180 mmHg, group C: SBP>180 mmHg), higher mean DeltaSBP (intra-arterial SBP-oscillometric SBP) levels were documented in group C (+19.8 mmHg, 95% confidence intervals: 12.2-27.4) when compared with groups B (+8.5 mmHg, 95% confidence intervals: 2.7-14.5; P=0.025) and A (+5.9 mmHg, 95% confidence intervals: 1.8-9.9; P=0.002).

Conclusion: Noninvasive automatic oscillometric BP measurements underestimate direct SBP recordings and overestimate direct DBP readings in acute stroke. The magnitude of the discrepancy between intra-arterial and oscillometric SBP recordings is even more prominent in patients with critically elevated SBP levels.

Citing Articles

Non-invasive versus arterial pressure monitoring in the pre-hospital critical care environment: a paired comparison of concurrently recorded measurements.

Perera Y, Raitt J, Poole K, Metcalfe D, Lewinsohn A Scand J Trauma Resusc Emerg Med. 2024; 32(1):77.

PMID: 39192296 PMC: 11348748. DOI: 10.1186/s13049-024-01240-y.


Comparison of intraoperative blood pressure values measured by noninvasive versus invasive methods during normotension, hypertension, and hypotension.

Irimpan J, Kesavan R, Rajan S, Kumar L J Anaesthesiol Clin Pharmacol. 2024; 40(2):258-263.

PMID: 38919432 PMC: 11196061. DOI: 10.4103/joacp.joacp_439_22.


Arterial Monitoring in Hypertensive Emergencies: Significance for the Critical Care Resuscitation Unit.

Tran Q, Gelmann D, Zahid M, Palmer J, Hollis G, Engelbrecht-Wiggans E West J Emerg Med. 2023; 24(4):763-773.

PMID: 37527376 PMC: 10393462. DOI: 10.5811/westjem.59373.


Most patients with non-hypertensive diseases at a critical care resuscitation unit require arterial pressure monitoring: a prospective observational study.

Engelbrecht-Wiggans E, Palmer J, Hollis G, Albelo F, Ali A, Hart E World J Emerg Med. 2023; 14(3):173-178.

PMID: 37152525 PMC: 10156524. DOI: 10.5847/wjem.j.1920-8642.2023.035.


Intra-beat biomarker for accurate continuous non-invasive blood pressure monitoring.

Abiri A, Chou E, Qian C, Rinehart J, Khine M Sci Rep. 2022; 12(1):16772.

PMID: 36202815 PMC: 9537243. DOI: 10.1038/s41598-022-19096-6.