Reliability and Validity of Non-invasive Determined Haemoglobin Mass and Blood Volumes

Overview
Date 2017 Jan 31
PMID 28135764
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: The carbon monoxide (CO) rebreathing method used for the determination of haemoglobin mass (Hb ) is associated with blood sample analysis (in this study: Radiometer ABL800). As an alternative hereto the aim of the present study was to evaluate the use of a portable and non-invasive CO pulse oximeter (Rad-57).

Method: With simultaneous determination of CO in the circulation by ABL800 (%HbCO) and Rad-57 (SpCO), Hb and blood volume (BV) were determined in duplicates in 24 volunteers. Percentage of typical errors (%TE) within methods and linear correlations between the two procedures were computed.

Results: Hb (Rad-57 = 798 ± 230 g; ABL800 = 781 ± 192 g) and BV (Rad-57 = 5700 ± 1373 ml; ABL800 = 5581 ± 1096 ml) were similar between methods. However, the %TE for Hb was higher (P<0·001) for Rad-57 (5·84 ± 5·29%) than for ABL800 (1·35 ± 1·13%). Similarly, the %TE for BV was higher (P<0·001) for Rad-57 (6·06 ± 5·76%) than for ABL800 (1·48 ± 1·25%). Lower (P<0·05) correlation coefficients between the methods were found when Hb  > 905 g and BV > 6193 ml.

Conclusion: Assessment of SpCO by Rad-57 resulted in considerably less precise determinations of Hb and BV, especially for high values. Thus, non-invasive assessment of Hb and BV cannot be recommended for scientific purposes, but may nonetheless be useful in clinical settings.

Citing Articles

Validation of a clinically applicable device for fast and accurate quantification of blood volume.

Andersen A, Bonne T, Hansen J, Oturai P, Lundby C J Clin Lab Anal. 2023; 37(9-10):e24928.

PMID: 37332175 PMC: 10388222. DOI: 10.1002/jcla.24928.


A semi-automated device rapidly determine circulating blood volume in healthy males and carbon monoxide uptake kinetics of arterial and venous blood.

Andersen A, Baungaard S, Bejder J, Graae J, Hristovska A, Agerskov M J Clin Monit Comput. 2022; 37(2):437-447.

PMID: 36201093 DOI: 10.1007/s10877-022-00921-2.


Baroreflex responses during dry resting and exercise apnoeas in air and pure oxygen.

Taboni A, Vinetti G, Fontolliet T, Grasso G, Tam E, Moia C Eur J Appl Physiol. 2020; 121(2):539-547.

PMID: 33151437 PMC: 7862076. DOI: 10.1007/s00421-020-04544-w.


Hematological Adaptations to Prolonged Heat Acclimation in Endurance-Trained Males.

Oberholzer L, Siebenmann C, Mikkelsen C, Junge N, Piil J, Morris N Front Physiol. 2019; 10:1379.

PMID: 31749713 PMC: 6842970. DOI: 10.3389/fphys.2019.01379.


Red blood cell volume is not decreased in ESA-naive anemic chronic kidney disease patients.

Lundby C, Ponte B, Lundby A, Robach P, de Seigneux S Physiol Rep. 2018; 6(21):e13900.

PMID: 30426716 PMC: 6234145. DOI: 10.14814/phy2.13900.