Performance Evaluation of the Touchscreen-based Muse™ Auto CD4/CD4% Single-platform System for CD4 T Cell Numeration in Absolute Number and in Percentage Using Blood Samples from Children and Adult Patients Living in the Central African Republic

Overview
Journal J Transl Med
Publisher Biomed Central
Date 2016 Nov 26
PMID 27884153
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The new microcapillary and fluorescence-based EC IVD-qualified Muse™ Auto CD4/CD4% single-platform assay (EMD Millipore Corporation, Merck Life Sciences, KGaA, Darmstadt, Germany) for CD4 T cell numeration in absolute number and in percentage was evaluated using Central African patients' samples compared against the reference EC IVD-qualified BD FACSCount (Becton-Dickinson, USA) flow cytometer.

Methods: EDTA-blood samples from 124 adults, 10 adolescents, 13 children and 3 infants were tested in parallel at 2 reference laboratories in Bangui.

Results: The Muse™ technique was highly reproducible, with low intra- and inter-run variabilities less than 15%. CD4 T cell counts of Muse™ and BD FACSCount in absolute number and percentage were highly correlated (r = 0.99 and 0.98, respectively). The mean absolute bias between Muse™ and BD FACSCount cells in absolute number and percentage were -5.91 cells/µl (95% CI -20.90 to 9.08) with limits of agreement from -77.50 to 202.40 cells/µl, and +1.69 %CD4 (95% CI ±1.29 to +2.09), respectively. The percentages of outliers outside the limits of agreement were nearly similar in absolute number (8%) and percentage (10%). CD4 T cell counting by Muse™ allowed identifying the majority of individuals with CD4 T cell <200, <350 or <750 cells/µl corresponding to the relevant thresholds of therapeutic care, with sensitivities of 95.5-100% and specificities of 83.9-100%.

Conclusions: The Muse™ Auto CD4/CD4% Assay analyzer is a reliable alternative flow cytometer for CD4 T lymphocyte enumeration to be used in routine immunological monitoring according to World Health Organization recommendations in HIV-infected adults as well as children living in resource-constrained settings.

Citing Articles

Sphingosine-1-phosphate Decreases Erythrocyte Dysfunction Induced by β-Amyloid.

Misiti F, Diotaiuti P, Lombardo G, Tellone E Int J Mol Sci. 2024; 25(10).

PMID: 38791223 PMC: 11121638. DOI: 10.3390/ijms25105184.


[Evaluation of the performances of the "MUSE AUTO CD4/CD4%" flow cytometer vs "GUAVA AUTO CD4/CD4%" flow cytometer for measuring the rate of CD4 lymphocytes in patients infected with HIV in Cameroon].

Kalla G, Voundi E, Tanghu F, Medi Sike C, Noubom M, Nkodo J Pan Afr Med J. 2019; 32:2.

PMID: 31068996 PMC: 6492305. DOI: 10.11604/pamj.2019.32.2.16990.


CD4-T cell enumeration in human immunodeficiency virus (HIV)-infected patients: A laboratory performance evaluation of Muse Auto CD4/CD4% system by World Health Organization prequalification of in vitro diagnostics.

Ceulemans A, Bouzahzah C, Prat I, Urassa W, Kestens L PLoS One. 2019; 14(1):e0209677.

PMID: 30673713 PMC: 6343963. DOI: 10.1371/journal.pone.0209677.


Diagnosis of Human Immunodeficiency Virus Infection.

Parekh B, Ou C, Fonjungo P, Kalou M, Rottinghaus E, Puren A Clin Microbiol Rev. 2018; 32(1).

PMID: 30487166 PMC: 6302353. DOI: 10.1128/CMR.00064-18.

References
1.
Phillips A, Pillay D, Miners A, Bennett D, Gilks C, Lundgren J . Outcomes from monitoring of patients on antiretroviral therapy in resource-limited settings with viral load, CD4 cell count, or clinical observation alone: a computer simulation model. Lancet. 2008; 371(9622):1443-51. DOI: 10.1016/S0140-6736(08)60624-8. View

2.
Babiker A, Gilks C, Munderi P, Kabuye G, Kasirye R, Zalwango E . Routine versus clinically driven laboratory monitoring of HIV antiretroviral therapy in Africa (DART): a randomised non-inferiority trial. Lancet. 2009; 375(9709):123-31. PMC: 2805723. DOI: 10.1016/S0140-6736(09)62067-5. View

3.
Dieye T, Diaw P, Daneau G, Wade D, Sylla Niang M, Camara M . Evaluation of a flow cytometry method for CD4 T cell enumeration based on volumetric primary CD4 gating using thermoresistant reagents. J Immunol Methods. 2011; 372(1-2):7-13. DOI: 10.1016/j.jim.2011.07.012. View

4.
Bendavid E, Young S, Katzenstein D, Bayoumi A, Sanders G, Owens D . Cost-effectiveness of HIV monitoring strategies in resource-limited settings: a southern African analysis. Arch Intern Med. 2008; 168(17):1910-8. PMC: 2894578. DOI: 10.1001/archinternmed.2008.1. View

5.
Bergeron M, Ding T, Elharti E, Oumzil H, Soucy N, Harmouche H . Evaluation of a dry format reagent alternative for CD4 T-cell enumeration for the FACSCount system: a report on a Moroccan-Canadian study. Cytometry B Clin Cytom. 2009; 78(3):188-93. DOI: 10.1002/cyto.b.20501. View