Toward a Robust Estimation of Respiratory Rate From Pulse Oximeters

Overview
Date 2016 Nov 23
PMID 27875128
Citations 45
Authors
Affiliations
Soon will be listed here.
Abstract

Goal: Current methods for estimating respiratory rate (RR) from the photoplethysmogram (PPG) typically fail to distinguish between periods of high- and low-quality input data, and fail to perform well on independent "validation" datasets. The lack of robustness of existing methods directly results in a lack of penetration of such systems into clinical practice. The present work proposes an alternative method to improve the robustness of the estimation of RR from the PPG.

Methods: The proposed algorithm is based on the use of multiple autoregressive models of different orders for determining the dominant respiratory frequency in the three respiratory-induced variations (frequency, amplitude, and intensity) derived from the PPG. The algorithm was tested on two different datasets comprising 95 eight-minute PPG recordings (in total) acquired from both children and adults in different clinical settings, and its performance using two window sizes (32 and 64 seconds) was compared with that of existing methods in the literature.

Results: The proposed method achieved comparable accuracy to existing methods in the literature, with mean absolute errors (median, 25[Formula: see text]-75[Formula: see text] percentiles for a window size of 32 seconds) of 1.5 (0.3-3.3) and 4.0 (1.8-5.5) breaths per minute (for each dataset respectively), whilst providing RR estimates for a greater proportion of windows (over 90% of the input data are kept).

Conclusion: Increased robustness of RR estimation by the proposed method was demonstrated.

Significance: This work demonstrates that the use of large publicly available datasets is essential for improving the robustness of wearable-monitoring algorithms for use in clinical practice.

Citing Articles

Parallel Datasets for Classification of Respiratory Rhythm Phases.

Szymanski J, Szefler M, Karski K, Krawczak F, Jankowski D Sci Data. 2025; 12(1):346.

PMID: 40011528 PMC: 11865572. DOI: 10.1038/s41597-025-04625-5.


The MSPTDfast photoplethysmography beat detection algorithm: design, benchmarking, and open-source distribution.

Charlton P, Arguello-Prada E, Arguello Prada E, Mant J, Kyriacou P, Kyriacou P Physiol Meas. 2025; 46(3).

PMID: 39978069 PMC: 11894679. DOI: 10.1088/1361-6579/adb89e.


Real-Time Anomaly Detection in Physiological Parameters: A Multi-Squad Monitoring and Communication Architecture.

Jinad R, Gupta K, Oladimeji D, Rasheed A, Varol C Sensors (Basel). 2025; 25(3).

PMID: 39943567 PMC: 11819872. DOI: 10.3390/s25030929.


Non-Intrusive Monitoring of Vital Signs in the Lower Limbs Using Optical Sensors.

Simoes J, Oliveira R, Costa F, Teixeira A, Leitao C, Correia P Sensors (Basel). 2025; 25(2).

PMID: 39860673 PMC: 11768218. DOI: 10.3390/s25020305.


LieRHRV system for remote lie detection using heart rate variability parameters.

Davoodi M, Aspis N, Drori Y, Weiser-Bitoun I, Yaniv Y Sci Rep. 2024; 14(1):30749.

PMID: 39730487 PMC: 11681120. DOI: 10.1038/s41598-024-80480-5.


References
1.
Verkruysse W, Svaasand L, Nelson J . Remote plethysmographic imaging using ambient light. Opt Express. 2008; 16(26):21434-45. PMC: 2717852. DOI: 10.1364/oe.16.021434. View

2.
Karlen W, Raman S, Ansermino J, Dumont G . Multiparameter respiratory rate estimation from the photoplethysmogram. IEEE Trans Biomed Eng. 2013; 60(7):1946-53. DOI: 10.1109/TBME.2013.2246160. View

3.
Orphanidou C, Bonnici T, Charlton P, Clifton D, Vallance D, Tarassenko L . Signal-quality indices for the electrocardiogram and photoplethysmogram: derivation and applications to wireless monitoring. IEEE J Biomed Health Inform. 2014; 19(3):832-8. DOI: 10.1109/JBHI.2014.2338351. View

4.
Larsen P, Harty M, Thiruchelvam M, Galletly D . Spectral analysis of AC and DC components of the pulse photoplethysmograph at rest and during induction of anaesthesia. Int J Clin Monit Comput. 1997; 14(2):89-95. DOI: 10.1007/BF03356582. View

5.
Poh M, McDuff D, Picard R . Advancements in noncontact, multiparameter physiological measurements using a webcam. IEEE Trans Biomed Eng. 2010; 58(1):7-11. DOI: 10.1109/TBME.2010.2086456. View