The Viewpoint-specific Failure of Modern 3D Displays in Laparoscopic Surgery

Overview
Specialty General Surgery
Date 2016 Aug 20
PMID 27539218
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Surgeons conventionally assume the optimal viewing position during 3D laparoscopic surgery and may not be aware of the potential hazards to team members positioned across different suboptimal viewing positions. The first aim of this study was to map the viewing positions within a standard operating theatre where individuals may experience visual ghosting (i.e. double vision images) from crosstalk. The second aim was to characterize the standard viewing positions adopted by instrument nurses and surgical assistants during laparoscopic pelvic surgery and report the associated levels of visual ghosting and discomfort.

Methods: In experiment 1, 15 participants viewed a laparoscopic 3D display from 176 different viewing positions around the screen. In experiment 2, 12 participants (randomly assigned to four clinically relevant viewing positions) viewed laparoscopic suturing in a simulation laboratory. In both experiments, we measured the intensity of visual ghosting. In experiment 2, participants also completed the Simulator Sickness Questionnaire.

Results: We mapped locations within the dimensions of a standard operating theatre at which visual ghosting may result during 3D laparoscopy. Head height relative to the bottom of the image and large horizontal eccentricities away from the surface normal were important contributors to high levels of visual ghosting. Conventional viewing positions adopted by instrument nurses yielded high levels of visual ghosting and severe discomfort.

Conclusions: The conventional viewing positions adopted by surgical team members during laparoscopic pelvic operations are suboptimal for viewing 3D laparoscopic displays, and even short periods of viewing can yield high levels of discomfort.

Citing Articles

Comparison of laparoscopic gastrectomy with 3-D/HD and 2-D/4 K camera system for gastric cancer: a prospective randomized control study.

Kanaji S, Yamazaki Y, Kudo T, Harada H, Takiguchi G, Urakawa N Langenbecks Arch Surg. 2021; 407(1):105-112.

PMID: 34458930 DOI: 10.1007/s00423-021-02302-w.


Laparoscopic skills training: the effects of viewing mode (2D vs. 3D) on skill acquisition and transfer.

Beattie K, Hill A, Horswill M, Grove P, Stevenson A Surg Endosc. 2020; 35(8):4332-4344.

PMID: 32876737 PMC: 8263422. DOI: 10.1007/s00464-020-07923-8.


Three-dimensional Two-dimensional Laparoscopic Surgery for Colorectal Cancer: Systematic Review and Meta-analysis.

Pantalos G, Patsouras D, Spartalis E, Dimitroulis D, Tsourouflis G, Nikiteas N In Vivo. 2019; 34(1):11-21.

PMID: 31882458 PMC: 6984079. DOI: 10.21873/invivo.11740.


The use of 3D laparoscopic imaging systems in surgery: EAES consensus development conference 2018.

Arezzo A, Vettoretto N, Francis N, Bonino M, Curtis N, Amparore D Surg Endosc. 2018; 33(10):3251-3274.

PMID: 30515610 DOI: 10.1007/s00464-018-06612-x.


[3 D laparoscopy versus 2 D laparoscopy : An up to date evaluation].

Buia A, Farkas S Chirurg. 2018; 89(10):777-784.

PMID: 30076442 DOI: 10.1007/s00104-018-0693-3.


References
1.
Sakata S, Grove P, Stevenson A, Hewett D . The impact of three-dimensional imaging on polyp detection during colonoscopy: a proof of concept study. Gut. 2016; 65(5):730-1. DOI: 10.1136/gutjnl-2016-311507. View

2.
Lee C, Lin H, Tsai C . Designs of broadband and wide-view patterned polarizers for stereoscopic 3D displays. Opt Express. 2011; 18(26):27079-94. DOI: 10.1364/OE.18.027079. View

3.
Kennedy R, Berbaum K, Lilienthal M . Disorientation and postural ataxia following flight simulation. Aviat Space Environ Med. 1997; 68(1):13-7. View

4.
Zaroff C, Knutelska M, Frumkes T . Variation in stereoacuity: normative description, fixation disparity, and the roles of aging and gender. Invest Ophthalmol Vis Sci. 2003; 44(2):891-900. DOI: 10.1167/iovs.02-0361. View

5.