» Articles » PMID: 27535684

Evaluation of the Aptima HIV-1 Quant Dx Assay Using Plasma and Dried Blood Spots

Overview
Specialty Microbiology
Date 2016 Aug 19
PMID 27535684
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

HIV-1 RNA quantitation in plasma, or virus load testing, is the primary method by which the response to antiretroviral therapy is monitored. Here we describe evaluation of the Aptima HIV-1 Quant Dx assay (Aptima) performed on the automated Panther system. The clinical performance of Aptima was compared to that of the Cobas AmpliPrep/Cobas TaqMan HIV-1 Test v2.0 (CAP/CTM) using 162 EDTA plasma samples collected from patients undergoing HIV-1 monitoring. Overall agreement was 84.0% (136/162), with a kappa statistic of 0.723 (standard error, 0.047; 95% confidence interval [CI], 0.630 to 0.815), indicating substantial agreement. Using the 86 clinical samples quantifiable by both methods, Passing-Bablok regression revealed a regression line of Y = (1.069 × X) - 0.346 (95% CI of the slope [1.003 to 1.139] and intercept [-0.666 to -0.074]), and Bland-Altman analysis demonstrated a mean difference (Aptima-CAP/CTM) of -0.075 log10 copies/ml (95% limits of agreement of -0.624 to 0.475), consistent with negative bias. Comparison of Aptima results for paired dried blood spot (DBS) and plasma specimens archived from participants in the Peninsula AIDS Research Cohort Study (PARC) demonstrated an overall agreement of 94.7% (90/95) when 1,000 copies/ml was used as the threshold. In conclusion, the Aptima HIV-1 Quant Dx assay provides a suitable alternative for HIV-1 monitoring in plasma and DBS.

Citing Articles

The performance of using dried blood spot specimens for HIV-1 viral load testing: A systematic review and meta-analysis.

Vojnov L, Carmona S, Zeh C, Markby J, Boeras D, Prescott M PLoS Med. 2022; 19(8):e1004076.

PMID: 35994520 PMC: 9447868. DOI: 10.1371/journal.pmed.1004076.


Comparison of the performance of Aptima HIV-1 Quant Dx Assay with Abbott RealTime HIV Assay for viral load monitoring using plasma and Dried Blood Spots collected in Kenya.

Mwau M, Schaffer S, Kimani H, Kasiano P, Ogolla F, Ajema E PLoS One. 2022; 17(8):e0269838.

PMID: 35994447 PMC: 9394820. DOI: 10.1371/journal.pone.0269838.


Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya.

Mwau M, Danda J, Mbugua J, Handa A, Fortunko J, Worlock A PLoS One. 2021; 16(4):e0249376.

PMID: 33798221 PMC: 8018649. DOI: 10.1371/journal.pone.0249376.


Performance evaluation of the Aptima HIV-1 RNA Quant assay on the Panther system using the standard and dilution protocols.

Rossetti R, Smith T, Luo W, Taussig J, Valentine-Graves M, Sullivan P J Clin Virol. 2020; 129:104479.

PMID: 32531665 PMC: 7395893. DOI: 10.1016/j.jcv.2020.104479.


Evaluation of the Aptima HCV Quant Dx Assay Using Serum and Dried Blood Spots.

Weber J, Sahoo M, Taylor N, Shi R, Pinsky B J Clin Microbiol. 2019; 57(4).

PMID: 30760534 PMC: 6440767. DOI: 10.1128/JCM.00030-19.


References
1.
Zida S, Tuaillon E, Barro M, Kwimatouo Lekpa Franchard A, Kagone T, Nacro B . Estimation of HIV-1 DNA Level Interfering with Reliability of HIV-1 RNA Quantification Performed on Dried Blood Spots Collected from Successfully Treated Patients. J Clin Microbiol. 2016; 54(6):1641-1643. PMC: 4879270. DOI: 10.1128/JCM.03372-15. View

2.
Smit P, Sollis K, Fiscus S, Ford N, Vitoria M, Essajee S . Systematic review of the use of dried blood spots for monitoring HIV viral load and for early infant diagnosis. PLoS One. 2014; 9(3):e86461. PMC: 3945725. DOI: 10.1371/journal.pone.0086461. View

3.
Xu S, Song A, Nie J, Li X, Meng S, Zhang C . Comparison between the automated Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 test version 2.0 assay and its version 1 and Nuclisens HIV-1 EasyQ version 2.0 assays when measuring diverse HIV-1 genotypes in China. J Clin Virol. 2011; 53(1):33-7. DOI: 10.1016/j.jcv.2011.10.001. View

4.
Pas S, Rossen J, Schoener D, Thamke D, Pettersson A, Babiel R . Performance evaluation of the new Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 test version 2.0 for quantification of human immunodeficiency virus type 1 RNA. J Clin Microbiol. 2010; 48(4):1195-200. PMC: 2849552. DOI: 10.1128/JCM.01832-09. View

5.
Garrett N, Apea V, Nori A, Ushiro-Lumb I, Oliver A, Baily G . Comparison of the rate and size of HIV-1 viral load blips with Roche COBAS TaqMan HIV-1 versions 1.0 and 2.0 and implications for patient management. J Clin Virol. 2012; 53(4):354-5. DOI: 10.1016/j.jcv.2011.12.024. View