Multicenter Evaluation of Two Next-Generation HIV-1 Quantitation Assays, Aptima Quant Dx and Cobas 6800, in Comparison to the RealTie HIV-1 Reference Assay

Overview
Specialty Microbiology
Date 2018 Aug 3
PMID 30068537
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

High accuracy and precision at the lower end of quantification are crucial requirements of a modern HIV viral load (VL) assay, since some clinically relevant thresholds are located at 50 and 200 copies/ml. In this study, we compared the performance of two new fully automated HIV-1 VL assays, Aptima HIV-1 Quant Dx and Cobas HIV-1 (Cobas 6800), with the established RealTie 2000 assay. Assay precision and accuracy were evaluated in a retrospective evaluation out of excess plasma material from four HIV-1+ individuals (subtypes B, C, CRF01_AE, and CRF02_AG). Native plasma samples were diluted to nominal concentrations at 50 and 200 copies/ml (according to the RealTie 2000 assay). All dilutions were tested in triplicate in five independent runs over 5 days and in three labs per system. Assay concordance was determined using 1,011 surplus clinical routine samples, as well as selected retrospective longitudinal samples from 7 patients on treatment. The three assays yielded highly concordant results for individual clinical samples ( > 0.98; average difference, ≤0.2 log copies/ml) and retrospective longitudinal samples from patients on treatment. The Aptima and RealTie assays showed similar high precision, meeting the 5σ criterion for the majority of samples across all labs and subtypes. The Cobas assay was less precise, missing the 5σ criterion for the majority of samples at low concentrations. In this analysis, results from the Cobas assay appeared less reliable near the clinically relevant cutoff and should be interpreted with more caution in this context. Due to high precision, full automation, and high concordance with the RealTie assay, the Aptima assay represents a good alternative in routine VL monitoring.

Citing Articles

The Perils of Overly Sensitive Viral Load Testing for Persons With Human Immunodeficiency Virus.

Rodriguez M, Syros A, Rodriguez A, Serota D Open Forum Infect Dis. 2023; 10(10):ofad494.

PMID: 37849507 PMC: 10578508. DOI: 10.1093/ofid/ofad494.


The plasma separation card as a novel solution for enhancing central laboratory capability for HIV-1 viral load monitoring in limited-access settings.

Kiyaga C, Makoha C, Nkugwa I, Okiira C, Okwir R, Gebreab S PLOS Glob Public Health. 2023; 3(6):e0002099.

PMID: 37379313 PMC: 10306227. DOI: 10.1371/journal.pgph.0002099.


Comparison of the performance of Aptima HIV-1 Quant Dx Assay with Abbott RealTime HIV Assay for viral load monitoring using plasma and Dried Blood Spots collected in Kenya.

Mwau M, Schaffer S, Kimani H, Kasiano P, Ogolla F, Ajema E PLoS One. 2022; 17(8):e0269838.

PMID: 35994447 PMC: 9394820. DOI: 10.1371/journal.pone.0269838.


Performance Evaluation of the Aptima Assays in Comparison with the cobas 6800 Assays for the Detection of HIV-1, HBV, and HCV in Clinical Samples.

Park Y, Roh J, Kim S Ann Lab Med. 2022; 42(4):447-456.

PMID: 35177565 PMC: 8859551. DOI: 10.3343/alm.2022.42.4.447.


HIV-1 viral blips are associated with repeated and increasingly high levels of cell-associated HIV-1 RNA transcriptional activity.

Suzuki K, Levert A, Yeung J, Starr M, Cameron J, Williams R AIDS. 2021; 35(13):2095-2103.

PMID: 34148986 PMC: 8505147. DOI: 10.1097/QAD.0000000000003001.


References
1.
Garrett N, Apea V, Nori A, Ushiro-Lumb I, Oliver A, Baily G . Comparison of the rate and size of HIV-1 viral load blips with Roche COBAS TaqMan HIV-1 versions 1.0 and 2.0 and implications for patient management. J Clin Virol. 2012; 53(4):354-5. DOI: 10.1016/j.jcv.2011.12.024. View

2.
Nair S, Kim H, Fortunko J, Foote T, Peling T, Tran C . Aptima HIV-1 Quant Dx--A fully automated assay for both diagnosis and quantification of HIV-1. J Clin Virol. 2016; 77:46-54. DOI: 10.1016/j.jcv.2016.02.002. View

3.
Ouma K, Basavaraju S, Okonji J, Williamson J, Thomas T, Mills L . Evaluation of quantification of HIV-1 RNA viral load in plasma and dried blood spots by use of the semiautomated Cobas Amplicor assay and the fully automated Cobas Ampliprep/TaqMan assay, version 2.0, in Kisumu, Kenya. J Clin Microbiol. 2013; 51(4):1208-18. PMC: 3666812. DOI: 10.1128/JCM.03048-12. View

4.
Kessler H, Stelzl E . Profile of Roche's cobas® HCV tests. Expert Rev Mol Diagn. 2017; 17(4):311-319. DOI: 10.1080/14737159.2017.1293526. View

5.
Schalasta G, Borner A, Speicher A, Enders M . Comparative evaluation of the Aptima HIV-1 Quant Dx assay and COBAS TaqMan HIV-1 v2.0 assay using the Roche High Pure System for the quantification of HIV-1 RNA in plasma. Clin Chem Lab Med. 2015; 54(3):493-9. DOI: 10.1515/cclm-2015-0522. View