Thyroid Surgery with a Harmonic Scalpel: an Experimental Study

Overview
Publisher Dove Medical Press
Date 2016 Jun 30
PMID 27354833
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The goal of the study was to determinate the safety of the harmonic scalpel, widely used in thyroidectomy, near the recurrent laryngeal nerve (RLN).

Methods: The study involved ten pigs of either sex. Twenty RLNs at risk were dissected using the new harmonic scalpel FOCUS. The distances between the nerve and the activated instrument were checked with a millimeter ruler. After dissection, the pigs were euthanized, and both RLNs were fixed in formol and examined by histology after staining with hematoxylin-eosin. Due to technical reasons, only 18 RLNs from the ten pigs could be examined.

Results: In the experiment that investigated the extent of heat injury, ultrasonic dissection did not cause any immediate damage of the nerve even close to the RLN (1 mm away from the RLN).

Conclusion: The use of harmonic scalpel FOCUS for thyroid surgery is safe for the surrounding structures (nerves). Careful tissue applications of the device near the RLN (1 mm) did not cause any lesion histologically.

Citing Articles

Defining the Thyroid-RLN Entry Triangle for Enhanced Recurrent Laryngeal Nerve Exposure in TOETVA: A Retrospective Study.

Zhou S, Li H, Wu P, Li W, Wang Z, Zhang L J Otolaryngol Head Neck Surg. 2024; 53:19160216241301328.

PMID: 39704391 PMC: 11662381. DOI: 10.1177/19160216241301328.


Safety and Cost-effectiveness of LigaSure® in Total Thyroidectomy in Comparison with Conventional Suture Tie Technique.

Bhettani M, Rehman M, Khan M, Altaf H, Hakeem Khan K, Farooqui F Cureus. 2020; 11(12):e6368.

PMID: 31938650 PMC: 6957032. DOI: 10.7759/cureus.6368.


Safety of energy based devices for hemostasis in thyroid surgery.

Dionigi G, Wu C, Kim H, Liu X, Liu R, Randolph G Gland Surg. 2016; 5(5):490-494.

PMID: 27867863 PMC: 5106374. DOI: 10.21037/gs.2016.09.01.

References
1.
Tessema B, Roark R, Pitman M, Weissbrod P, Sharma S, Schaefer S . Observations of recurrent laryngeal nerve injury and recovery using a rat model. Laryngoscope. 2009; 119(8):1644-51. DOI: 10.1002/lary.20293. View

2.
Kinoshita T, Kanehira E, Omura K, Kawakami K, Watanabe Y . Experimental study on heat production by a 23.5-kHz ultrasonically activated device for endoscopic surgery. Surg Endosc. 1999; 13(6):621-5. DOI: 10.1007/s004649901055. View

3.
Carlander J, Johansson K, Lindstrom S, Velin A, Jiang C, Nordborg C . Comparison of experimental nerve injury caused by ultrasonically activated scalpel and electrosurgery. Br J Surg. 2005; 92(6):772-7. DOI: 10.1002/bjs.4948. View

4.
Jiang H, Shen H, Jiang D, Zheng X, Zhang W, Lu L . Evaluating the safety of the Harmonic Scalpel around the recurrent laryngeal nerve. ANZ J Surg. 2010; 80(11):822-6. DOI: 10.1111/j.1445-2197.2010.05436.x. View

5.
Quah H, Jayne D, Eu K, Seow-Choen F . Bladder and sexual dysfunction following laparoscopically assisted and conventional open mesorectal resection for cancer. Br J Surg. 2002; 89(12):1551-6. DOI: 10.1046/j.1365-2168.2002.02275.x. View