Experimental Study on Heat Production by a 23.5-kHz Ultrasonically Activated Device for Endoscopic Surgery

Overview
Journal Surg Endosc
Publisher Springer
Date 1999 May 29
PMID 10347305
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

An experimental study was carried out to evaluate heat production by an ultrasonically activated device (USAD) using an animal model. In an anesthetized living pig, the gastroepiploic and mesenteric vessels were coagulated and cut by an USAD at a power level of 70% (n = 8) or 100% (n = 8). During the division, the time-discrete temperature change on the surface of the animal tissue adjacent to the blade was measured by thermography. To compare the USAD with conventional electrocautery (EC), a full-thickness incision of the gastric wall was performed by each device, and the temperature change was measured. With the USAD, the temperature increased gradually and remained below 150 degrees C during the entire activating time at both power levels. By contrast, with EC at 30 W, the temperature increased rapidly and exceeded 350 degrees C within only a few seconds. The area above 60 degrees C reached a final width of 10 mm for the USAD, as compared with 22 mm for EC. Microscopically, thermal alterations such as carbonization and vaporization were much more severe and extensive in the adjacent tissue when using EC rather than the USAD. With the USAD, heat production is much slower and more limited than with conventional EC; thus, the USAD causes fewer thermal alterations in adjacent tissue. USAD should be preferred for tissue coagulation and cutting during endscopic surgery.

Citing Articles

Is endourological intervention a suitable treatment option in the management of iatrogenic thermal ureteral injury? A contemporary case series.

Cebeci O BMC Urol. 2022; 22(1):137.

PMID: 36057579 PMC: 9441037. DOI: 10.1186/s12894-022-01094-5.


Improving Voice Outcomes After Thyroid Surgery - Review of Safety Parameters for Using Energy-Based Devices Near the Recurrent Laryngeal Nerve.

Wang J, Huang T, Wu C, Lin Y, Tseng H, Liu C Front Endocrinol (Lausanne). 2021; 12:793431.

PMID: 34899616 PMC: 8662988. DOI: 10.3389/fendo.2021.793431.


Use of ultrasonic scalpel and monopolar electrocautery for skin incisions in neck dissection: a prospective randomized trial.

Schneider D, Goppold K, Kaemmerer P, Schoen G, Woehlke M, Bschorer R Oral Maxillofac Surg. 2018; 22(2):169-175.

PMID: 29492789 DOI: 10.1007/s10006-018-0686-x.


Thyroid surgery with a harmonic scalpel: an experimental study.

Dequanter D, Lammens M, Nagy N, Shahla M, Deniz Y, Aubert C Med Devices (Auckl). 2016; 9:139-42.

PMID: 27354833 PMC: 4910611. DOI: 10.2147/MDER.S104786.


Effect of metal surface characteristics on the adhesion performance of the integrated low-level energies method of adhesion.

Aodai T, Masuzawa T, Ozeki K, Kishida A, Higami T J Artif Organs. 2012; 15(4):386-94.

PMID: 22933053 DOI: 10.1007/s10047-012-0656-2.