Fast Blood Flow Monitoring in Deep Tissues with Real-time Software Correlators

Overview
Specialty Radiology
Date 2016 May 28
PMID 27231588
Citations 62
Authors
Affiliations
Soon will be listed here.
Abstract

We introduce, validate and demonstrate a new software correlator for high-speed measurement of blood flow in deep tissues based on diffuse correlation spectroscopy (DCS). The software correlator scheme employs standard PC-based data acquisition boards to measure temporal intensity autocorrelation functions continuously at 50 - 100 Hz, the fastest blood flow measurements reported with DCS to date. The data streams, obtained in vivo for typical source-detector separations of 2.5 cm, easily resolve pulsatile heart-beat fluctuations in blood flow which were previously considered to be noise. We employ the device to separate tissue blood flow from tissue absorption/scattering dynamics and thereby show that the origin of the pulsatile DCS signal is primarily flow, and we monitor cerebral autoregulation dynamics in healthy volunteers more accurately than with traditional instrumentation as a result of increased data acquisition rates. Finally, we characterize measurement signal-to-noise ratio and identify count rate and averaging parameters needed for optimal performance.

Citing Articles

An affordable, wearable, fiber-free pulsed-mode diffuse speckle contrast flowmetry (PM-DSCF) sensor for noninvasive measurements of deep cerebral blood flow.

Yeo C, Liu X, Mohtasebi M, Akbari F, Fathi F, Yu G ArXiv. 2025; .

PMID: 39990801 PMC: 11844631.


Proof of concept validation of bioresorbable optical fibers for diffuse correlation spectroscopy.

Pandayil J, Boetti N, Janner D, Durduran T, Cortese L Biomed Opt Express. 2024; 15(11):6384-6398.

PMID: 39553874 PMC: 11563325. DOI: 10.1364/BOE.540137.


ATLAS: a large array, on-chip compute SPAD camera for multispeckle diffuse correlation spectroscopy.

Gorman A, Finlayson N, Erdogan A, Fisher L, Wang Y, Mattioli Della Rocca F Biomed Opt Express. 2024; 15(11):6499-6515.

PMID: 39553854 PMC: 11563329. DOI: 10.1364/BOE.531416.


Modified Beer-Lambert algorithm to measure pulsatile blood flow, critical closing pressure, and intracranial hypertension.

Baker W, Forti R, Heye P, Heye K, Lynch J, Yodh A Biomed Opt Express. 2024; 15(9):5511-5532.

PMID: 39296411 PMC: 11407241. DOI: 10.1364/BOE.529150.


A Device-on-Chip Solution for Real-Time Diffuse Correlation Spectroscopy Using FPGA.

Moore C, Sunar U, Lin W Biosensors (Basel). 2024; 14(8).

PMID: 39194613 PMC: 11352433. DOI: 10.3390/bios14080384.


References
1.
Baker W, Parthasarathy A, Ko T, Busch D, Abramson K, Tzeng S . Pressure modulation algorithm to separate cerebral hemodynamic signals from extracerebral artifacts. Neurophotonics. 2015; 2(3):035004. PMC: 4524732. DOI: 10.1117/1.NPh.2.3.035004. View

2.
Baker W, Parthasarathy A, Busch D, Mesquita R, Greenberg J, Yodh A . Modified Beer-Lambert law for blood flow. Biomed Opt Express. 2014; 5(11):4053-75. PMC: 4242038. DOI: 10.1364/BOE.5.004053. View

3.
Zhou C, Yu G, Furuya D, Greenberg J, Yodh A, Durduran T . Diffuse optical correlation tomography of cerebral blood flow during cortical spreading depression in rat brain. Opt Express. 2009; 14(3):1125-44. DOI: 10.1364/oe.14.001125. View

4.
Jaillon F, Li J, Dietsche G, Elbert T, Gisler T . Activity of the human visual cortex measured non-invasively by diffusing-wave spectroscopy. Opt Express. 2009; 15(11):6643-50. DOI: 10.1364/oe.15.006643. View

5.
Choe R, Putt M, Carlile P, Durduran T, Giammarco J, Busch D . Optically measured microvascular blood flow contrast of malignant breast tumors. PLoS One. 2014; 9(6):e99683. PMC: 4072684. DOI: 10.1371/journal.pone.0099683. View