» Articles » PMID: 27230934

First Results and Follow-up of a Second-generation Circular Mapping and Ablation Catheter

Overview
Publisher Springer
Date 2016 May 28
PMID 27230934
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Pulmonary vein isolation with radiofrequency energy is widely used as a strategy for catheter ablation of atrial fibrillation (AF). Anatomically designed catheters have been developed to increase the efficiency of AF ablation procedures. The second-generation circular ablation catheter, PVAC GOLD, was re-designed to improve energy delivery and mitigate emboli. We investigated the procedural efficiency, biophysics, and chronic efficacy of PVAC GOLD in patients with AF.

Methods: We consecutively enrolled 40 patients (60 ± 11 years) with highly symptomatic, drug refractory AF. The first 20 patients were treated with the first-generation PVAC. The subsequent 20 patients were treated with the second-generation PVAC GOLD catheter. All patients were followed up at 3, 6, and 12 months.

Results: All 164 targeted PVs were successfully isolated. Ablations performed with PVAC GOLD showed a significant reduction in total number of ablations needed for PVI, fluoroscopy, and procedure times compared to PVAC (34.7 ± 7.0 vs. 27.0 ± 6.5; p = 0.009), fluoroscopy (29.5 ± 9.5 vs. 23.4 ± 7.0; p = 0.026), and procedure time (93.8 ± 18.9 vs. 83.1 ± 10.6; p = 0.033). PVAC GOLD showed improved biophysics including a reduction of low power ablations and an increase in mean effective energy delivery. At 12 months follow-up, AF recurrence rates were comparable in the two groups (35 vs. 30 %; p = 0.735). There were no adverse events.

Conclusions: The redesigned PVAC GOLD catheter demonstrates a reduction in radiofrequency ablation and procedure time and improved biophysics while maintaining chronic efficacy compared to the first-generation PVAC.

Citing Articles

Invasive strategies for rhythm control of atrial fibrillation: a narrative review.

Kim H, Lee C J Yeungnam Med Sci. 2024; 41(4):279-287.

PMID: 39307574 PMC: 11534407. DOI: 10.12701/jyms.2024.00703.


Atrial fibrillation: Primary prevention, secondary prevention, and prevention of thromboembolic complications: Part 2.

Trohman R, Huang H, Sharma P Front Cardiovasc Med. 2023; 9:1060096.

PMID: 36969508 PMC: 10036779. DOI: 10.3389/fcvm.2022.1060096.


Pulmonary vein isolation using second-generation single-shot devices: not all the same?.

Seidl P, Steinborn F, Costello-Boerrigter L, Surber R, Schulze P, Bottcher C J Interv Card Electrophysiol. 2020; 60(3):521-528.

PMID: 32415555 DOI: 10.1007/s10840-020-00751-9.


The Impact of Advances in Atrial Fibrillation Ablation Devices on the Incidence and Prevention of Complications.

Kece F, Zeppenfeld K, Trines S Arrhythm Electrophysiol Rev. 2018; 7(3):169-180.

PMID: 30416730 PMC: 6141930. DOI: 10.15420/aer.2018.7.3.


Clinical outcomes of AF patients treated with the first and second-generation of circular mapping and ablation catheter: insights from a real world multicenter experience.

Rovaris G, De Filippo P, Laurenzi F, Zanotto G, Bottoni N, Pozzi M J Interv Card Electrophysiol. 2017; 50(3):245-251.

PMID: 28875448 DOI: 10.1007/s10840-017-0278-y.


References
1.
Lewalter T, Weiss C, Spencker S, Jung W, Haverkamp W, Willems S . Gold vs. platinum-iridium tip catheter for cavotricuspid isthmus ablation: the AURUM 8 study. Europace. 2010; 13(1):102-8. DOI: 10.1093/europace/euq339. View

2.
Gal P, Aarntzen A, Smit J, Adiyaman A, Ramdat Misier A, Delnoy P . Conventional radiofrequency catheter ablation compared to multi-electrode ablation for atrial fibrillation. Int J Cardiol. 2014; 176(3):891-5. DOI: 10.1016/j.ijcard.2014.08.034. View

3.
Haines D, Stewart M, Dahlberg S, Barka N, Condie C, Fiedler G . Microembolism and catheter ablation I: a comparison of irrigated radiofrequency and multielectrode-phased radiofrequency catheter ablation of pulmonary vein ostia. Circ Arrhythm Electrophysiol. 2013; 6(1):16-22. DOI: 10.1161/CIRCEP.111.973453. View

4.
von Bary C, Weber S, Dornia C, Eissnert C, Fellner C, Latzin P . Evaluation of pulmonary vein stenosis after pulmonary vein isolation using a novel circular mapping and ablation catheter (PVAC). Circ Arrhythm Electrophysiol. 2011; 4(5):630-6. DOI: 10.1161/CIRCEP.111.963397. View

5.
Wieczorek M, Hoeltgen R, Akin E, Salili A, Oral H, Morady F . Results of short-term and long-term pulmonary vein isolation for paroxysmal atrial fibrillation using duty-cycled bipolar and unipolar radiofrequency energy. J Cardiovasc Electrophysiol. 2009; 21(4):399-405. DOI: 10.1111/j.1540-8167.2009.01640.x. View