Technical Report: Reference Photon Dosimetry Data for Varian Accelerators Based on IROC-Houston Site Visit Data

Overview
Journal Med Phys
Specialty Biophysics
Date 2016 May 6
PMID 27147349
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Accurate data regarding linear accelerator (Linac) radiation characteristics are important for treatment planning system modeling as well as regular quality assurance of the machine. The Imaging and Radiation Oncology Core-Houston (IROC-H) has measured the dosimetric characteristics of numerous machines through their on-site dosimetry review protocols. Photon data are presented and can be used as a secondary check of acquired values, as a means to verify commissioning a new machine, or in preparation for an IROC-H site visit.

Methods: Photon data from IROC-H on-site reviews from 2000 to 2014 were compiled and analyzed. Specifically, data from approximately 500 Varian machines were analyzed. Each dataset consisted of point measurements of several dosimetric parameters at various locations in a water phantom to assess the percentage depth dose, jaw output factors, multileaf collimator small field output factors, off-axis factors, and wedge factors. The data were analyzed by energy and parameter, with similarly performing machine models being assimilated into classes. Common statistical metrics are presented for each machine class. Measurement data were compared against other reference data where applicable.

Results: Distributions of the parameter data were shown to be robust and derive from a student's t distribution. Based on statistical and clinical criteria, all machine models were able to be classified into two or three classes for each energy, except for 6 MV for which there were eight classes. Quantitative analysis of the measurements for 6, 10, 15, and 18 MV photon beams is presented for each parameter; supplementary material has also been made available which contains further statistical information.

Conclusions: IROC-H has collected numerous data on Varian Linacs and the results of photon measurements from the past 15 years are presented. The data can be used as a comparison check of a physicist's acquired values. Acquired values that are well outside the expected distribution should be verified by the physicist to identify whether the measurements are valid. Comparison of values to this reference data provides a redundant check to help prevent gross dosimetric treatment errors.

Citing Articles

PRIMO Monte Carlo software as a tool for commissioning of an external beam radiotherapy treatment planning system.

Calvo-Ortega J, Hermida-Lopez M Rep Pract Oncol Radiother. 2023; 28(4):529-540.

PMID: 37795225 PMC: 10547427. DOI: 10.5603/RPOR.a2023.0060.


Characterizing the interplay of treatment parameters and complexity and their impact on performance on an IROC IMRT phantom using machine learning.

Mehrens H, Molineu A, Hernandez N, Court L, Howell R, Jaffray D Radiother Oncol. 2023; 182:109577.

PMID: 36841341 PMC: 10121814. DOI: 10.1016/j.radonc.2023.109577.


AAPM MEDICAL PHYSICS PRACTICE GUIDELINE 5.b: Commissioning and QA of treatment planning dose calculations-Megavoltage photon and electron beams.

Geurts M, Jacqmin D, Jones L, Kry S, Mihailidis D, Ohrt J J Appl Clin Med Phys. 2022; 23(9):e13641.

PMID: 35950259 PMC: 9512346. DOI: 10.1002/acm2.13641.


The current status and shortcomings of stereotactic radiosurgery.

Mehrens H, Nguyen T, Edward S, Hartzell S, Glenn M, Branco D Neurooncol Adv. 2022; 4(1):vdac058.

PMID: 35664554 PMC: 9154323. DOI: 10.1093/noajnl/vdac058.


Photon beam modeling variations predict errors in IMRT dosimetry audits.

Glenn M, Brooks F, Peterson C, Howell R, Followill D, Pollard-Larkin J Radiother Oncol. 2021; 166:8-14.

PMID: 34748857 PMC: 8863621. DOI: 10.1016/j.radonc.2021.10.021.


References
1.
Sjostrom D, Bjelkengren U, Ottosson W, Behrens C . A beam-matching concept for medical linear accelerators. Acta Oncol. 2008; 48(2):192-200. DOI: 10.1080/02841860802258794. View

2.
Watts R . Comparative measurements on a series of accelerators by the same vendor. Med Phys. 2000; 26(12):2581-5. DOI: 10.1118/1.598796. View

3.
Cho S, Vassiliev O, Lee S, Liu H, Ibbott G, Mohan R . Reference photon dosimetry data and reference phase space data for the 6 MV photon beam from varian clinac 2100 series linear accelerators. Med Phys. 2005; 32(1):137-48. DOI: 10.1118/1.1829172. View

4.
Ikoro N, Johnson D, Antich P . Characteristics of the 6-MV photon beam produced by a dual energy linear accelerator. Med Phys. 1987; 14(1):93-7. DOI: 10.1118/1.596113. View

5.
Fontenla D, Napoli J, Chui C . Beam characteristics of a new model of 6-MV linear accelerator. Med Phys. 1992; 19(2):343-9. DOI: 10.1118/1.596864. View