Meta-analytic Estimation of Measurement Variability and Assessment of Its Impact on Decision-making: the Case of Perioperative Haemoglobin Concentration Monitoring

Overview
Publisher Biomed Central
Date 2016 Jan 21
PMID 26787309
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Background: As a part of a larger Health Technology Assessment (HTA), the measurement error of a device used to monitor the hemoglobin concentration of a patient undergoing surgery, as well as its decision consequences, were to be estimated from published data.

Methods: A Bayesian hierarchical model of measurement error, allowing the meta-analytic estimation of both central and dispersion parameters (under the assumption of normality of measurement errors) is proposed and applied to published data; the resulting potential decision errors are deduced from this estimation. The same method is used to assess the impact of an initial calibration.

Results: The posterior distributions are summarized as mean ± sd (credible interval). The fitted model exhibits a modest mean expected error (0.24 ± 0.73 (-1.23 1.59) g/dL) and a large variability (mean absolute expected error 1.18 ± 0.92 (0.05 3.36) g/dL). The initial calibration modifies the bias (-0.20 ± 0.87 (-1.99 1.49) g/dL), but the variability remains almost as large (mean absolute expected error 1.05 ± 0.87 (0.04 3.21) g/dL). This entails a potential decision error ("false positive" or "false negative") for about one patient out of seven.

Conclusions: The proposed hierarchical model allows the estimation of the variability from published aggregates, and allows the modeling of the consequences of this variability in terms of decision errors. For the device under assessment, these potential decision errors are clinically problematic.

Citing Articles

Current Status of Measurement Accuracy for Total Hemoglobin Concentration in the Clinical Context.

Stawschenko E, Schaller T, Kern B, Bode B, Dorries F, Kusche-Vihrog K Biosensors (Basel). 2022; 12(12).

PMID: 36551114 PMC: 9775510. DOI: 10.3390/bios12121147.


Diagnostic accuracy of point-of-care devices for detection of anemia in community settings in India.

Neogi S, Sharma J, Pandey S, Zaidi N, Bhattacharya M, Kar R BMC Health Serv Res. 2020; 20(1):468.

PMID: 32456639 PMC: 7249358. DOI: 10.1186/s12913-020-05329-9.


Meta-analysis of variance: an illustration comparing the effects of two dietary interventions on variability in weight.

Senior A, Gosby A, Lu J, Simpson S, Raubenheimer D Evol Med Public Health. 2016; 2016(1):244-55.

PMID: 27491895 PMC: 4981479. DOI: 10.1093/emph/eow020.

References
1.
Kim S, Choi J, Kim H, Choi S, Choi I . Continuous noninvasive hemoglobin measurement is useful in patients undergoing double-jaw surgery. J Oral Maxillofac Surg. 2014; 72(9):1813-9. DOI: 10.1016/j.joms.2014.03.011. View

2.
Dewhirst E, Naguib A, Winch P, Rice J, Galantowicz M, McConnell P . Accuracy of noninvasive and continuous hemoglobin measurement by pulse co-oximetry during preoperative phlebotomy. J Intensive Care Med. 2013; 29(4):238-42. DOI: 10.1177/0885066613485355. View

3.
Kim S, Lilot M, Murphy L, Sidhu K, Yu Z, Rinehart J . Accuracy of continuous noninvasive hemoglobin monitoring: a systematic review and meta-analysis. Anesth Analg. 2014; 119(2):332-346. DOI: 10.1213/ANE.0000000000000272. View

4.
Park Y, Lee J, Song H, Byon H, Kim H, Kim J . The accuracy of noninvasive hemoglobin monitoring using the radical-7 pulse CO-Oximeter in children undergoing neurosurgery. Anesth Analg. 2012; 115(6):1302-7. DOI: 10.1213/ANE.0b013e31826b7e38. View

5.
Frasca D, Mounios H, Giraud B, Boisson M, Debaene B, Mimoz O . Continuous monitoring of haemoglobin concentration after in-vivo adjustment in patients undergoing surgery with blood loss. Anaesthesia. 2015; 70(7):803-9. DOI: 10.1111/anae.13028. View