Validity of the Polar V800 Heart Rate Monitor to Measure RR Intervals at Rest

Overview
Specialty Physiology
Date 2015 Dec 29
PMID 26708360
Citations 129
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To assess the validity of RR intervals and short-term heart rate variability (HRV) data obtained from the Polar V800 heart rate monitor, in comparison to an electrocardiograph (ECG).

Method: Twenty participants completed an active orthostatic test using the V800 and ECG. An improved method for the identification and correction of RR intervals was employed prior to HRV analysis. Agreement of the data was assessed using intra-class correlation coefficients (ICC), Bland-Altman limits of agreement (LoA), and effect size (ES).

Results: A small number of errors were detected between ECG and Polar RR signal, with a combined error rate of 0.086 %. The RR intervals from ECG to V800 were significantly different, but with small ES for both supine corrected and standing corrected data (ES <0.001). The bias (LoA) were 0.06 (-4.33 to 4.45 ms) and 0.59 (-1.70 to 2.87 ms) for supine and standing intervals, respectively. The ICC was >0.999 for both supine and standing corrected intervals. When analysed with the same HRV software no significant differences were observed in any HRV parameters, for either supine or standing; the data displayed small bias and tight LoA, strong ICC (>0.99) and small ES (≤0.029).

Conclusions: The V800 improves over previous Polar models, with narrower LoA, stronger ICC and smaller ES for both the RR intervals and HRV parameters. The findings support the validity of the Polar V800 and its ability to produce RR interval recordings consistent with an ECG. In addition, HRV parameters derived from these recordings are also highly comparable.

Citing Articles

Test-Retest Reliability and Concurrent Validity of Photoplethysmography Finger Sensor to Collect Measures of Heart Rate Variability.

Rogers D, Himariotis A, Sherriff T, Proulx Q, Duong M, Noel S Sports (Basel). 2025; 13(2).

PMID: 39997960 PMC: 11861371. DOI: 10.3390/sports13020029.


Enhancing Psychophysiological Well-Being Through Nature-Based Soundscapes: An Examination of Heart Rate Variability in a Cross-Over Study.

Kumpulainen S, Esmaeilzadeh S, Pesonen M, Brazao C, Pesola A Psychophysiology. 2025; 62(1):e14760.

PMID: 39803887 PMC: 11726612. DOI: 10.1111/psyp.14760.


Exploring the Dynamics of Canine-Assisted Interactions: A Wearable Approach to Understanding Interspecies Well-Being.

Holder T, Nichols C, Summers E, Roberts D, Bozkurt A Animals (Basel). 2025; 14(24).

PMID: 39765532 PMC: 11672835. DOI: 10.3390/ani14243628.


Diagnostic accuracy of heart rate variability as a screening tool for mild neurocognitive disorder.

Czopek-Rowinska J, de Bruin E, Manser P Front Aging Neurosci. 2025; 16:1498687.

PMID: 39741522 PMC: 11685156. DOI: 10.3389/fnagi.2024.1498687.


Comparison of 1-hour floatation-REST versus conventional napping on heart rate variability in active individuals.

Besson C, De Stefani G, Baggish A, Schmitt L, Millet G, Gremeaux V BMJ Open Sport Exerc Med. 2024; 11(1):e002292.

PMID: 39691209 PMC: 11647359. DOI: 10.1136/bmjsem-2024-002292.


References
1.
Gamelin F, Berthoin S, Bosquet L . Validity of the polar S810 heart rate monitor to measure R-R intervals at rest. Med Sci Sports Exerc. 2006; 38(5):887-93. DOI: 10.1249/01.mss.0000218135.79476.9c. View

2.
Weir J . Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005; 19(1):231-40. DOI: 10.1519/15184.1. View

3.
Nunan D, Jakovljevic D, Donovan G, Hodges L, Sandercock G, Brodie D . Levels of agreement for RR intervals and short-term heart rate variability obtained from the Polar S810 and an alternative system. Eur J Appl Physiol. 2008; 103(5):529-37. DOI: 10.1007/s00421-008-0742-6. View

4.
Vanderlei L, Silva R, Pastre C, Azevedo F, Godoy M . Comparison of the Polar S810i monitor and the ECG for the analysis of heart rate variability in the time and frequency domains. Braz J Med Biol Res. 2008; 41(10):854-9. DOI: 10.1590/s0100-879x2008005000039. View

5.
Nunan D, Donovan G, Jakovljevic D, Hodges L, Sandercock G, Brodie D . Validity and reliability of short-term heart-rate variability from the Polar S810. Med Sci Sports Exerc. 2008; 41(1):243-50. DOI: 10.1249/MSS.0b013e318184a4b1. View