In Vivo Subjective and Objective Longitudinal Chromatic Aberration After Bilateral Implantation of the Same Design of Hydrophobic and Hydrophilic Intraocular Lenses
Overview
Affiliations
Purpose: To measure the longitudinal chromatic aberration in vivo using psychophysical and wavefront-sensing methods in patients with bilateral implantation of monofocal intraocular lenses (IOLs) of similar aspheric design but different materials (hydrophobic Podeye and hydrophilic Poday).
Setting: Instituto de Optica, Consejo Superior de Investigaciones Cientificas, Madrid, Spain.
Design: Prospective observational study.
Methods: Measurements were performed with the use of psychophysical (480 to 700 nm) and wavefront-sensing (480 to 950 nm) methods using a custom-developed adaptive optics system. Chromatic difference-of-focus curves were obtained from best-focus data at each wavelength, and the longitudinal chromatic aberration was obtained from the slope of linear regressions to those curves.
Results: The longitudinal chromatic aberration from psychophysical measurements was 1.37 diopters (D) ± 0.08 (SD) (hydrophobic) and 1.21 ± 0.08 D (hydrophilic). From wavefront-sensing, the longitudinal chromatic aberration was 0.88 ± 0.07 D and 0.73 ± 0.09 D, respectively. At 480 to 950 nm, the longitudinal chromatic aberration was 1.27 ± 0.09 D (hydrophobic) and 1.02 ± 0.13 D (hydrophilic). The longitudinal chromatic aberration was consistently higher in eyes with the hydrophobic IOL than in eyes with the hydrophilic IOL (a difference of 0.16 D and 0.15 D, respectively). Similar to findings in young phakic eyes, the longitudinal chromatic aberration from the psychophysical method was consistently higher than from wavefront-sensing, by 0.48 D (35.41%) for the hydrophobic IOL and 0.48 D (39.43%) for the hydrophilic IOL.
Conclusion: Longitudinal chromatic aberrations were smaller with hydrophilic IOLs than with hydrophobic IOLs of the same design.
Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned.
How do intraocular lens materials influence the outcome of cataract surgery?.
Grzybowski A, Auffarth G, LaHood B Curr Opin Ophthalmol. 2024; 36(1):18-24.
PMID: 39446645 PMC: 11620326. DOI: 10.1097/ICU.0000000000001095.
Labuz G, Yan W, Baur I, Khoramnia R, Auffarth G Sci Rep. 2023; 13(1):14781.
PMID: 37679352 PMC: 10484925. DOI: 10.1038/s41598-023-41634-z.
Bellucci C, Mora P, Tedesco S, Gandolfi S, Bellucci R Ophthalmol Ther. 2023; 12(2):1387-1395.
PMID: 36602719 PMC: 9815056. DOI: 10.1007/s40123-022-00646-0.
Adaptive optics visual simulators: a review of recent optical designs and applications [Invited].
Marcos S, Artal P, Atchison D, Hampson K, Legras R, Lundstrom L Biomed Opt Express. 2023; 13(12):6508-6532.
PMID: 36589577 PMC: 9774875. DOI: 10.1364/BOE.473458.
Vision is protected against blue defocus.
Benedi-Garcia C, Vinas M, Dorronsoro C, Burns S, Peli E, Marcos S Sci Rep. 2021; 11(1):352.
PMID: 33432060 PMC: 7801416. DOI: 10.1038/s41598-020-79911-w.